Troubleshooting
54
HORIBA
●
Part of the display is missing
If part of the display is missing, it is the Instrument defect. Contact your dealer.
It is possible to check part of the display is missing. The display can be set to full screen
display. Follow the steps below to check the display.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
3. Use the
keys to select the "LCD" (full screen display) and then press the
key.
The display changes to full screen display. Check if part of the display is missing.
To return to the setting mode, press the
key.
key
key
key