Maintenance and storage
55
●
Maintenance and storage of electrodes
This section describes an overview of the procedures for maintenance and storage of pH, ORP
and DO electrodes.
●
How to clean the pH and ORP electrodes
Always clean the electrode with deionized water after every measurement. When the response
is slow or residue from the sample adheres to the electrode, use the appropriate method below
to clean the electrode, and then clean again with deionized water.
For pH electrode
For ORP electrode
●
How to clean the membrane of DO electrode
The membrane of DO electrode is extremely thin. Take care, not to tear the membrane when
cleaning. Clean the membrane with deionized water and wipe it with a soft cloth, take care not
to damage it.
When using a neutral cleaning solution to clean the DO electrode, keep the neutral cleaning
solution away from the membrane in order to prevent deterioration of the membrane.
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Oil
Alcohol, or diluted neutral cleaning solution
Inorganic substance
1 mol/L HCl or electrode cleaning solution
Protein
Cleaning solution including protein-removing enzyme
Alkali
Dip in 1 mol/L HCl or electrode cleaning solution for 1h to 2 h
Type of dirt
Cleaning solution
General
Dilute neutral cleaning solution (General dish washing liquid
works reasonably well.)
Oil
Inorganic substance
Immerse dilute nitric acid (1:1 nitric acid)
Note
Содержание LAQUA-PD210
Страница 1: ......
Страница 2: ......
Страница 49: ...Data set up 35 DO mode...
Страница 51: ...Data set up 37 DO mode...
Страница 53: ...Data set up 39 DO mode...
Страница 56: ...General set up 42 DO mode...
Страница 58: ...General set up 44 DO mode...
Страница 60: ...General set up 46 DO mode...
Страница 62: ...General set up 48 DO mode...
Страница 65: ...Clock set up 51 DO mode...
Страница 67: ...Clock set up 53 DO mode...
Страница 86: ...Specifications 67...
Страница 88: ......