ORP calibration
13
●
ORP/mV Calibration
Calibration is necessary for accurate ORP measurement. To perform ORP calibration, follow
the procedure detailed below:
Prerequisites
• Clean the ORP electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the PD meter and plug in the ORP electrode.
• Prepare buffer solution required for calibration.
• Ensure that the meter is in mV measurement mode.
• Dip the ORP electrode into the standard solution ensuring that the solution level is at least
3 cm from the electrode tip.
• Absolute value measurement mode and relative value measurement mode are the two
types of measurement mode available for ORP (mV) measurement.
• In absolute value measurement mode, the handheld meter displays the actual voltage
value.
• In relative value measurement mode, user can adjust the absolute mV value by calibration.
If the mV value is adjusted, the meter automatically indicates relative mV value as
RmV
.
The adjustment mV is applied as an offset to the absolute mV value.
• In the relative mV mode, the absolute mV value can be adjusted by ± 200 mV.
To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip
Содержание LAQUA-PD210
Страница 1: ......
Страница 2: ......
Страница 49: ...Data set up 35 DO mode...
Страница 51: ...Data set up 37 DO mode...
Страница 53: ...Data set up 39 DO mode...
Страница 56: ...General set up 42 DO mode...
Страница 58: ...General set up 44 DO mode...
Страница 60: ...General set up 46 DO mode...
Страница 62: ...General set up 48 DO mode...
Страница 65: ...Clock set up 51 DO mode...
Страница 67: ...Clock set up 53 DO mode...
Страница 86: ...Specifications 67...
Страница 88: ......