Set the meter as shown below.
2
Prepare the standard solutions.
Prepare the two types of standard solution to create Cl
-
concentration ratio
1:10.
For example, if the sample solution concentration is 250 mg/L, prepare
standard solutions of 100 mg/L and 1000 mg/L. This case is used as an
example in the following steps.
3
Prepare the first solution for calibration.
(1) Put 50 mL of 100 mg/L standard solution into a 100 mL beaker.
(2) Add 1 mL of ISA.
(3) Stir the solution slowly with a magnetic stirrer.
Set the valence as shown in the
table on the right.
Setting item
Setting value
Valence -1
Always use new standard solution for calibration.
Note
Note
(1) Standard solution (100 mg/L)
(2) ISA
Beaker: 100 mL
(3)
Stir the solution slowly
with a magnetic stirrer.
1 mL
50 mL