Rectal Use: 

Lubricate the probe cover with a water-soluble jelly for easier 

insertion; do not use petroleum jelly, as it may cause an inaccurate reading. 
Gently insert the tip of the probe no more than 1/2 inch into the rectum. 

 

STOP

 if you meet any resistance. The sensing unit is on  the very tip of the probe 

and there is no need to insert the probe deep into the rectum. 

Dispose of the used probe cover and wash the thermometer as recommended. 
The peak temperature should be reached in approximately 25-35 seconds. 

 

Axillary Use: 

Wipe armpit with a dry towel. Place the probe in the 

patient’s armpit (Figure 3) and keep the patient’s arm pressed 
firmly against the body. In order to achieve a better axillary reading, 
a minimum measuring time of 90 seconds or longer is 
recommended, depending on placement of the probe and
closure of the arm against the body. 

 

Battery Replacement

 

1.   When the symbol “      ” appears, the battery  needs to 

     be replaced. 

2.   Remove battery compartment lid (Figure 4). 

3.   Use a pointed object such as a toothpick to pry loose the 

     old battery. Avoid using any sharp metal object. 

4.   Place new 1.55V coin-type battery (LR41 or equivalent) in the chamber with the 

     positive (+) side facing up or towards the display of the unit. 

5.   Replace the battery compartment lid. 

 

Cleaning and Disinfecting

 

       

Do not use corrosive detergent to clean the unit.

 

       

Clean and disinfect the thermometer before and after each use.

 

Cleaning Process

 

1.  Wipe probe with a cloth dipped in a disinfectant solution such as 70% 
     isopropyl (rubbing) alcohol or soap and water. 

2.  Rinse off disinfectant residue. 

 

Care and Storage

 

•   Do NOT drop or crush; this device is not shock proof. 
•   Do NOT dismantle or modify device other than replacing the battery. 

•   Do NOT store the device in an extreme environment with direct sunlight or high 

temperatures. 

•   Store the unit in the protective case while not in use. 

LIMITED WARRANTY

 

This thermometer is guaranteed for a period of 24 months from date of purchase. This warranty 
does not cover batteries, or damage resulting from improper handling or use. Specific legal rights 
stipulated in your state may guide this warranty. If you have a legitimate claim under this warranty, 
send item(s) postage paid to Service Center, Attn: Repair Dept., 6850 Southbelt Dr., Caledonia, MI 
49316. Please include your name and address, phone no., proof of purchase, and a brief note ex-
plaining the problem, your return address and $2.00 to cover shipping and handling. 

 
 

 

IB p/n 93-410HOP-00 rev 6 (4/25/2020)

Printed in China

LATEX

Figure 4

Specifications

 

       

Size:                            

5”L x 3/4”W x 3/8”H  (12.7cm L x 1.9 cm W x .95 cm H) 

       

Weight:                        

0.388 oz. (11 grams) 

       

Display:                       

LCD, Liquid Crystal Display 

       

Battery:                       

1.55V (LR41)  

       

Battery Life:                

Estimated 4,500 measurements 

       

Measurement Time:    

Oral 30* seconds / Rectal 25-35 seconds / Axillary 90 seconds 

       

Measurement Range:   

90.0°F to 109.9°F  (32.0°C to 42.9°C) 

                                              96.0°F to 107.0 °F   

±

0.2°F / 1°C               (34.0°C to 42.0°C) 

                                                 at a room temperature of 64.4°F to 82.4°F    (18.0°C to 28.0°C) 

       

Accuracy:

                     90.0°F to 96.0°F      

±

0.4°F / 0.2°C            (32.0°C to 34.0°C) 

                                                 and 107.0°F to 109.9°F                             (42.0°C to 42.9°C) 

       

Measurement Mode:   

Real measurement with peak-hold method 

       

Audible Alarm:             

1 beep at start-up. Continuous beep at measurement completion 

       

Auto Turn-off:              

In approximately 8 ±1 minutes 

       

Memory:                      

Last reading 

       

Operating Conditions:

  

50°F to 104°F  (10°C to 40°C)  

                                              Humidity of 15% to 95% (non-condensing) 

       

Storage Conditions:

    

-13°F to 131°F  (-25°C to 55°C)  

                                              Humidity of 15% to 95% (non-condensing) 

       

Standards:                   

Complies with EN 12470-3, ASTM E 1112-00, 

                                           IEC 60601-1, IEC 60601-1- 2, IEC 60601-1-11 
                                           Electromagnetic compatibility: device fulfills the stipulations 
                                           of standard IEC 60601-1-2 
                                           

The manufacturer recommends a calibration check every 2 years.

 

                                          

*30 second response time achieved in water bath test.

BATTERY

Service Center

 

6850 Southbelt Dr. 

Caledonia, MI 49316 

 

Inspected in the U.S.A. 

Made in China

Read the instruction manual carefully before using 
this device, especially the safety instructions, and 
keep the instruction manual for future use.

Type BF applied part

Batteries and electronic devices must be disposed of 
in accordance with the locally applicable                   
regulations, not with domestic waste.