Temperature test

 

A temperature test is required on diaphragm gas 

meters with temperature conversion BK..Te only.

 

The accuracy of the temperature measurement 

can be verified with this test.

 

The temperature test can only be carried out in 

Service mode.

 CAUTION

To avoid damage to the unit:

–   

Comply with ambient temperature, see 

page  15 (Technical data). Deviations from 
the permitted ambient temperature will be re-
corded in the error memory.

 

Temperature measurement accuracy, see 

page  15 (Technical data).

  

 Install the diaphragm gas meter in a climatic 

chamber.

  

 Activate Service mode – see page 5 (Service 

mode).

  

 Change to the “Temperature” menu.

 

The current gas temperature is displayed.

 4 

 Close the climatic chamber.

 5 

 Select an ambient temperature as a reference 

value and bring the climatic chamber to this tem-
perature.

 

To ensure there is also a uniform temperature 

in the meter, we recommend starting the meter 
air/gas flow during the temperature adjustment 
phase.

 

Ensure that temperature distribution remains 
uniform and stable during the temperature 
measurement.

 6 

  Compare the measured value to the temperature 

reference value.

 

If required, several reference values can be 
checked. In this case, repeat the test as of 
point 

 5 

.

Assistance in the event of 
malfunction

 ?  Fault
 !  Cause
 •  Remedy

Possible faults and suggested solutions
 ? 

 

The 

 symbol is displayed.

 ! 

 If the 

 symbol appears next to a measured 

value, this means that the value is invalid.

 • 

  After the next data synchronization, the data are 

recorded again correctly and 

 disappears.

 ? 

 

When pressing the user keys, the backlight-
ing and/or display remain switched off. A 
beep can nevertheless be heard.

 ! 

  Energy-saving mode is active. Due to excessive 

use of the index, the average energy consump-
tion has been exceeded.

 • 

  Leave the index unused for an extended period, 

e.g. 24 hours. After this, the user interface will 
once again be available.

 ? 

 

When pressing the user keys, the display 
remains switched off and no beep can be 
heard.

 ! 

 The index is defective.

 • 

 Contact the manufacturer.

 ? 

 

The 

 symbol is displayed.

 ! 

  Low battery. This symbol is only displayed when 

battery power is low.

 • 

 Replace the battery.

 

In the case of faults which are not described here, 
contact the manufacturer immediately.

Spare parts

Battery

For meters without communication module:
Elster Part No.: 32906520.
For meters with communication module: 
Elster Part No.: 72910321.

Communication module

Use ATEX compliant communication modules only. 

A safety sticker with the following code must be 

attached to the communication module:  

 II 3G Ex ic IIA Gc (-25°C ≤ Ta ≤ +55°C).

Electrical connection parameters, see page  15 
(Technical data).

Screw locking cap

Elster Part No.: 32447510.

Содержание Themis alpha EI2

Страница 1: ...TION Indicates possible material damage All interventions may only be carried out by qualified gas technicians Electrical interventions may only be carried out by qualified electricians Conversion spare parts All technical changes are prohibited Only use OEM spare parts Changes to edition 12 13 The following chapters have been changed Checking the usage Logistics Contents D GB GB 1 1000106001 002 ...

Страница 2: ...al number BK G E S N EI2 Qmax Qmin V pmax class 1 5 class 1 5 Valve Ve tm tg M 0102 12345678 ATEX The electronic index is suitable for use in potentially explosive atmospheres For the exact use zone see ATEX sticker on the diaphragm gas meter or see the operating instructions for diaphragm gas meters BK G1 6 to BK G25 http docuthek kromschroeder com doclib main php language 1 folderid 400041 by_cl...

Страница 3: ...ill automatically change back to the main screen when no user key has been pressed for 30 s and switches off after a further 30 s You can navigate from the main screen to the various menus such as Meter information us ing the user keys Menu overview The display can differ depending on the parameteriza tion or communication module Main menu Valve release procedure Current meter reading Icon definit...

Страница 4: ...played with date and time for the start and end of the period The meter reading is displayed for the start and end of the period in m3 The consumption for this period is indicated in m3 The tariff band may be displayed Electronic index with communication module The symbol is displayed if for example the tolerance between the internal time recording and the actual time is too large This can lead to...

Страница 5: ... process until all the pixels are off and Test instructions appears in the menu area Service mode is activated Service mode menu overview Test instructions Temperature optional Date time LCD pixel test Cyclic test UMI peripherals UMI 1 UMI 2 UMI 3 Main menu Optional contents from communication module Battery diagnosis Meter information Test instructions ON Test instructions Automatic return to mai...

Страница 6: ...atus Change required Battery diagnosis U hlc 3 60 V U min 3 00 V Status Test in progress Time 05 00 h m Battery diagnosis U hlc 3 60 V U min 3 00 V Status Removed Date and time Information on the date and time display Date and time 10 01 2011 10 02 06 The operator can transfer the switchover be tween winter and summer time to the commu nication module provided that it supports this The date is giv...

Страница 7: ...ommunication is enabled for 1 minute If the optical communications link is not used during this time the interface will be deactivated 3 Initiate communication The procedure depends on your user software For further information refer to the operating instructions for communication modules type ECM for electronic index Changing the battery WARNING Risk of explosion in explosion hazard areas As a ge...

Страница 8: ...s must be confirmed at the meter on site by the consumer or another person if applicable 1 Establish the optical communications link see page 7 Establishing an optical communica tions link 2 Release the valve The procedure depends on your user software ON Valve open safety check CHECK IF APPLIANCES OFF HOLD FOR GAS Opening valve Please wait Gas flow check complete Gas flow check successful The tes...

Страница 9: ...ructions or the servicing steps provided by the relevant operator 6 Replace the service cover 7 Push in a new screw locking cap see page 14 Spare parts The body carrying out this task should apply its own seal Replacing the communication module 1 Establish the optical communications link see page 7 Establishing an optical communica tions link 2 Deactivate the respective communication mod ule BK G ...

Страница 10: ...n the unit under test in K tb base temperature in C Vb15 converted volume with reference to tb 15 C Vb converted volume with reference to tb There are two different options for carrying out the cyclic test Option 1 cyclic test at a constant flow rate The test rig is in pre trial operation i e start of measurement on the unit under test will be de layed Maintain the flow rate constant Test load and...

Страница 11: ...olume VP the following applies TP 273 15 tg K where tg relevant gas temperature on the unit under test in C display The curly brackets mean numerical value of On a nozzle test rig with a known flow rate steps 2 and 6 can be omitted The error calculation is based on PTB Testing Instructions Volume 29 Messgeräte für Gas Gaszähler Measuring instruments for gas gas meters Edition 2003 Option 2 cyclic ...

Страница 12: ...tions link has been established The procedure depends on your user software 1 Establish the optical communications link see page 7 Establishing an optical communica tions link 2 Interrupt communication to the installed commu nication modules before starting the test so that the measurement accuracy will not be adversely affected As of step 3 continue as described in one of the two following sectio...

Страница 13: ...nder test VP VP N x VImp N number of pulses VImp pulse value in m 9 Read off the volume on the unit under test The inherent error of the master meter is to be taken into account if applicable Vact N VN 1 FN 100 10 Determine the measuring deviation on the unit under test FP 100 x VP x pP x TN Vact N x pN x TP 1 The pressure and temperature values of the unit under test corrected with reference to t...

Страница 14: ...f point 5 Assistance in the event of malfunction Fault Cause Remedy Possible faults and suggested solutions The symbol is displayed If the symbol appears next to a measured value this means that the value is invalid After the next data synchronization the data are recorded again correctly and disappears When pressing the user keys the backlight ing and or display remain switched off A beep can nev...

Страница 15: ... is certified as part of the electronic index Communication modules Please contact the manufacturer for details on suit able communication modules Information on the ATEX suitability can be found in the operating in structions of the respective communication module The UMI interface features the following parameters UO 3 9 V peak 3 6 V nominal IO 7 5 A peak 506 mA nominal PO 1 55 W CO 12 µF LO 50 ...

Страница 16: ...ide Stafford ST16 3EF Tel 44 1785 275200 Fax 44 1785 275305 jeavons info gb elster com www elstermetering co uk Ireland Active Energy Control Ltd Unit 4 Clare Marts Quin Road Ennis Co Clare Tel 353 65 6840600 Fax 353 65 6840610 info aec ie www aec ie Germany Elster GmbH Strotheweg 1 49504 Lotte Tel 49 541 1214 0 Fax 49 541 1214 370 info elster instromet com www elster instromet com Contact ...

Отзывы: