background image

 

 

SmartLine RMA801 HART/DE Remote Indicator Assembly 
Quick Start Installation Guide

 

34-ST-25-63, Revision 2, March 2019 

This document provides descriptions 
and procedures for the Quick 
Installation of Honeywell’s family of 
SmartLine Remote Indicator. 

The SmartLine Remote Indicator is a 
configurable intelligent field device 
which functions as an output and 
status indicator for any HART and 
DE devices. 

For full details refer to the manuals 
listed below for protocols, user 
Interface (HMI) operation, Installation, 
configuration, calibration, 
maintenance, parts, and safety and 
approvals etc. including options 

Copyrights, Notices and 
Trademarks.  
Copyright 2019 by Honeywell 
Revision 2, March 2019 
 
Trademarks 
SmartLine, RMA are U.S. registered 
trademarks of Honeywell Inc.  
 
HART® is Trademarks of 
FieldComm Group™ 

 

 

Table of Contents 

Installation and setup .................................................................................1 
Site evaluation............................................................................................1 

Installation precautions ..............................................................................1 

Explosion-Proof Conduit Seal ....................................................................1 
Mounting Remote indicator ........................................................................2 
Mounting Dimensions ................................................................................2 
Bracket Mounting .......................................................................................2 
Wiring a Remote Indicator .........................................................................2 
Terminal Block ...........................................................................................2 

Wiring Connections and Power Up ............................................................3 
Wiring procedure Options ..........................................................................3 

DEVICE CONFIGURATION ......................................................................3 
Appendix A. PRODUCT CERTIFICATIONS .............................................3

 

 

Figure 1: Electronics Housing Components ..............................................1 
Figure 2: Typical Pipe Mounted Installations .............................................2 
Figure 3: Mounting Bracket Secured to a Remote indicator .....................2 
Figure 4: Pipe Mounting Bracket Secured to a Horizontal or Vertical 
Pipe ............................................................................................................2 
Figure 5: Remote Indicator Secured to a Wall Mounting Bracket .............2 
Figure 6: DE/ANALOG Terminal Block ......................................................2 
Figure 7: RMA801 Terminal Block .............................................................2 
Figure 8: Remote Indicator connected to the negative loop wire ..............3 
Figure 9: Remote Indicator connected to the negative loop wire ..............3 
Figure 10: Remote Indicator connected to the positive loop wire .............3 

 

 

Documentation 

To access complete documentation, including language variants, scan  
the QR code below using your smart phone/device or QR code scanner. 

Go to the APP store for your free Smartphone QR scanner 

Or you can follow the URL to access the online SmartLine HUB page.  

The HUB page will contain direct links to open SmartLine product 
documentation. 

URL                                               QR Code

 

https://hwll.co/SmartLineHUB

 

 

 
 

 

 

   

 

Features and Options 

The RMA801 Remote Indicator provides a means of remote-mounting a indicator (display) 
that is associated with a Honeywell Smartline Transmitter or any transmitter operating in a 
4-20 mA current loop. 

The RMA801 is a DE/Analog Remote Indicator which can be connected anywhere along 
the current loop. 

For analog PV, the RMA801 measures the loop current and displays the equivalent PV 
value on the display.  
The RMA801 will autoconfigure when connected to Honeywell STT700/STT250 DE 
transmitters when a database upload is performed.  

 This document provides the information for a quick setup. For detailed information, please refer 
RMA801 user manual, 34-ST-25-62. 

 

 

Figure 1: Electronics Housing Components 

 CAUTION

 

Temperature  extremes  can  affect  display  quality.  The  display  can  go  blank  if  the 
temperature is below -20°C or above 70°C; however, this is only a temporary condition. The 
display will again be readable when temperatures return to within operable limits. 
 
The device shall be operated by a trained professional. It is the us

er/installer’s responsibility 

to  install  the  indicator  in  accordance  with  national  and  local  code  requirements.  Conduit 
entry  plugs  and  adapters  shall  be  suitable  for  the  environment,  shall  be  certified  for  the 
hazardous location when required and acceptable to the authority having jurisdiction for the 
plant  

The RMA device is always connected in series with the transmitter. The current loop will 
be broken if the RMA801 device is removed from the loop. 

 

 

I

nstallation and setup 

 

Site evaluation 

Evaluate the site selected for the Remote Indicator installation with respect to the 
process system design specifications and Honeywell’s published performance 
characteristics for your particular model. Some parameters that you may want to 
include in your site evaluation can be found in the RMA801 user manual, 34-ST-25-
62 

Installation precautions 

Temperature extremes can affect display quality. The display can go blank if the 
temperature is below -20°C or above +70°C; however, this is only a temporary 
condition. The display will again be readable when temperatures return to within 
operable limits. 

 
 
 
 
 

 

Explosion-Proof Conduit Seal 

 

WARNING

 

When installed as explosion proof in a Division 1 Hazardous Location, keep 
covers tight while the Remote Indicator is energized. Disconnect power to the 
Remote Indicator in the non-hazardous area prior to removing end caps for 
service. 

When installed as non-incendive equipment in a Division 2 hazardous location, 
disconnect power to the Remote Indicator in the non-hazardous area, or 
determine that the location is non-hazardous before disconnecting or 
connecting the Remote Indicator wires.

 

 
 
 

 

 

Содержание SmartLine RMA801

Страница 1: ...perating in a 4 20 mA current loop The RMA801 is a DE Analog Remote Indicator which can be connected anywhere along the current loop For analog PV the RMA801 measures the loop current and displays the equivalent PV value on the display The RMA801 will autoconfigure when connected to Honeywell STT700 STT250 DE transmitters when a database upload is performed This document provides the information f...

Страница 2: ...a Remote indicator Figure 4 Pipe Mounting Bracket Secured to a Horizontal or Vertical Pipe 4 Wall Mount Option Refer Figure 5 Position the bracket on the mounting surface at the desired location and secure the bracket to the mounting surface using the appropriate hardware Wall mounting hardware requirements to be determined and supplied by the end user Figure 5 Remote Indicator Secured to a Wall M...

Страница 3: ...h ground Connect the Loop Power shield to earth ground at the power supply end and not at RMA end The shield shall be continuous from transmitter through RMA till the power supply DEVICE CONFIGURATION Once the loop is powered the RMA will be turned ON with a scrolling message SmartLine appearing on the display For DE devices the RMA is autoconfigured For any DE configuration change in the transmit...

Страница 4: ...to 85o C Enclosure Type IP66 IP67 ALL ALL STANDARDS EN 60079 0 2012 A11 2013 EN 60079 1 2014 EN 60079 11 2012 EN 60079 31 2014 EN 60079 7 2015 D IECEx Flame proof Ex db IIC T6 T4 Gb Ex tb IIIC T 95o C Db Note 1 T6 50ºC to 65ºC T4 T5 50 ºC to 85ºC Intrinsically Safe Ex ia IIC T4 Ga Note 2 50o C to 70o C Non Incendive Ex ec IIC T4 Gc Note 1 50o C to 70o C Enclosure IP66 IP67 ALL ALL STANDARDS IEC 60...

Страница 5: ...LOSION HAZARD SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I DIVISION 2 Intrinsically Safe Divisions 1 Zone 1 and Zone 2 Environments WARNING EXPLOSION HAZARD SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY WARNING DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE IS PRESENT Explosion Proof Division 1 and Zone 1 Environments WARNING DO NOT OPEN WHEN AN EXPLOSIVE GAS ATMOSPHERE IS ...

Страница 6: ...44 0 1202645583 FAX 44 0 1344 655554 Email Sales sc cp apps salespa62 honeywell com or TAC hfs tac support honeywell com AMERICAS Honeywell Process Solutions Phone 1 800 423 9883 or 1 215 641 3610 TAC hfs tac support honeywell com Sales 1 800 343 0228 Email Sales FP Sales Apps honeywell com Knowledge Base search engine http bit ly 2N5Vldi WARRANTY REMEDY Honeywell warrants goods of its manufacture...

Отзывы: