![Honeywell SmartLine RMA801 Скачать руководство пользователя страница 53](http://html.mh-extra.com/html/honeywell/smartline-rma801/smartline-rma801_user-manual_151889053.webp)
Revision 2
RMA803 Remote Indicator User's Guide
42
Hazardous Locations Certifications
MSG
CODE
AGENCY
TYPE OF PROTECTION
Electrical
Parameters
Ambient
Temperature
C
ATEX
Flame-proof and Dust:
II 2 G Ex db IIC T6..T4 Gb
II 2 D Ex tb IIIC T 95
o
C Db
Note 1
T6: -50ºC to +65ºC
T4, T5: -50 ºC to
85ºC
Intrinsically Safe:
II 1 G Ex ia IIC T4 Ga
Note 2
-50
o
C to 70
o
C
Non-Incendive
II 3 G Ex ec IIC T4 Gc
Note 1
-50
o
C to 85
o
C
Enclosure:
Type IP66/ IP67
ALL
ALL
STANDARDS:
EN 60079-0: 2012+A11: 2013; EN 60079-1: 2014; EN 60079-11:
2012; EN 60079-31: 2014; EN 60079-7: 2015;
D
IECEx
Flame-proof:
Ex db IIC T6..T4 Gb
Ex tb IIIC T 95
o
C Db
Note 1
T6: -50ºC to +65ºC
T4, T5: -50 ºC to
85ºC
Intrinsically Safe:
Ex ia IIC T4 Ga
Note 2
-50
o
C to 70
o
C
Non-Incendive
Ex ec IIC T4 Gc
Note 1
-50
o
C to 70
o
C
Enclosure:
IP66/ IP67
ALL
ALL
STANDARDS:
IEC
60079-0: 2011; IEC 60079-1: 2014;
IEC 60079-11: 2011; IEC 60079-7: 2015; IEC 60079-31: 2014
K
CSA-Canada
Explosion proof:
Class I, Division 1, Groups A, B, C, D;T6..T4
Dust Ignition Proof:
Class II, III, Division 1, Groups E, F, G; T4
Class I Zone 1 Ex db IIC T4 Gb
Ex db IIC T4 Gb
Zone 21 Ex tb IIIC T 95
o
C Db
Ex tb IIIC T 95
o
C Db
Note 1
T6: -50ºC to +65ºC
T4, T5: -50 ºC to
85ºC
Intrinsically Safe:
CSA 14.2689056
Class I, II, III, Division 1, Groups A, B, C, D,
E, F, G; T4
Ex ia IIC T4 Ga
Note 2
-50
o
C to 70
o
C
Non-Incendive
Class I, Division 2, Groups A, B, C, D; T4
Class I Zone 2 Ex nA IIC T4 Gc
Ex nA IIC T4 Gc
Note 1
-50
o
C to 85
o
C
Enclosure:
Type 4X/ IP66/ IP67
ALL
ALL
Содержание SmartLine RMA801
Страница 1: ...Honeywell Process Solutions RMA801 Remote Indicator User s Guide 34 ST 03 62 Revision 2 0 March 2019 ...
Страница 2: ......
Страница 11: ......
Страница 50: ...39 RMA803 Remote Indicator User s Guide Revision 2 ...
Страница 51: ...Revision 2 RMA803 Remote Indicator User s Guide 40 ...
Страница 52: ...41 RMA803 Remote Indicator User s Guide Revision 2 ...
Страница 57: ...Revision 2 RMA803 Remote Indicator User s Guide 46 Control Drawing 50089981 ...
Страница 58: ...47 RMA803 Remote Indicator User s Guide Revision 2 ...
Страница 59: ...Revision 2 RMA803 Remote Indicator User s Guide 48 ...
Страница 60: ...49 RMA803 Remote Indicator User s Guide Revision 2 This page has been intentionally left blank ...