S301M CALIBRATION PROCEDURE  

 

510181 S301M _Calibration_Procedure_R02 

www.honeywellanalytics.com 

2

 

7.  To calibrate the device, scroll to Set Span** and change the span gas calibration value using this 

procedure;

  

 

Calib 

007

S301M

Status :

In calib...

Set Zero
Set Span

246

PPM

 

a.  Using the right arrow, move the cursor to xxx PPM (span value field).  Press Enter to select the 

field (it’s editable when flashing). 

b.  Use the up or down arrows to increase or decrease the value, press Enter to validate the new 

value. 

c.  Move the cursor back to Set Span and press Enter to start the calibration.   

 

The device Span is being calibrated.  The screen will display the device’s status as “In calib...” until the 
calibration is complete. 
 
*The device must be configured in the 301C’s database in order to be included in the device addresses 
displayed on screen. 
**When selecting Set Span, make sure that the device has been exposed to the appropriate calibration 
gas. 

Table I - Calibration Information 

 

 

Gas 

Scale 

Warm-up 

Flow Rate

 

 

 

 

C

3

H

8

 

0 – 50.1% LEL 

15 min. 

500 ml/min 

CO 

0 - 255 ppm 

15 min 

500 ml/min 

NO

2

 

0 – 10.2 ppm 

15 min. 

500 ml/min 

Comb 

0 – 102% LEL 

15 min. 

500 ml/min 

 

 

Figure I – Calibration equipment setup

 

 

 

Calibration Kit 

 

Standard

 Calibration Kit (58L or 103L cylinder)  

Order Number 

Description 

SKS301MCOQ1  

STANDARD CAL KIT FOR S301M CO, 103L (CO 200 ppm BAL N2) 

SKS301MNO2Q1  

STANDARD CAL KIT FOR S301M NO2, 58L (NO2 5 ppm BAL N2) 

SKS301MC3H8Q1  

STANDARD CAL KIT FOR S301M C3H8, 103L (C3H8 20% LEL BAL AIR) 

SKS301MCOMBQ1  

STANDARD CAL KIT FOR S301M COMBUSTIBLE, 103L (CH4 20% LEL BAL AIR) 

 
 
 

Содержание S301M

Страница 1: ... Scroll through the menu options and select option 7 Network 3 Scroll through the Network options screens to the last screen and select the Calibration option The Calibration screen contains four 4 lines of information Calib 007 S301M Status Normal Set Zero Set Span 246 PPM Line 1 Indicates the mode Calib meaning calibration the Modbus address of the device to calibrate 007 and the type of device ...

Страница 2: ...calib until the calibration is complete The device must be configured in the 301C s database in order to be included in the device addresses displayed on screen When selecting Set Span make sure that the device has been exposed to the appropriate calibration gas Table I Calibration Information Gas Scale Warm up Flow Rate C3H8 0 50 1 LEL 15 min 500 ml min CO 0 255 ppm 15 min 500 ml min NO2 0 10 2 p...

Страница 3: ...menu et sélectionner l option 7 Reseau 3 Dérouler les options de Reseau et sélectionner Etalonnage au dernier écran L écran d étalonnage contient quatre 4 lignes d information Calib 007 S301M Statut Normale Etal Zero Etal Plage 246 PPM Ligne 1 Indique le mode Calib l adresse de l unité pour l étalonnage 007 et le type d unité à étalonner S301M Ligne 2 Indique le statut Normale ou En Etalon de l un...

Страница 4: ...rre L écran affiche le statut de l unité En etalon jusqu à la terminaison de l étalonnage L unité doit être configurée dans la base de données du 301C pour figurer dans la liste d adresse d unité affichée à l écran Lors de la sélection d Etal Plage assurer que l unité a été exposée au gaz d étalonnage approprié Tableau II Information d étalonnage Gaz Plage Réchauffement Débit C3H8 0 50 1 LIE 15 mi...

Отзывы: