S301M CALIBRATION PROCEDURE  

 

1

 

www.honeywellanalytics.com 510181 

S301M_Calibration_Procedure_R02

 

The unit requires minimum warm-up time of fifteen (15) minutes before being put into service and 
calibrated.  Calibration of this unit must be completed using the 301M (see 301M Calibration Procedure) or 
using the 301C controller menus (see instructions below).

 

 

CONNECTING THE HARDWARE 

 

1.  Plug the calibration adaptor onto the gas sensor inlet on the unit.   
2.  Screw the regulator to the appropriate calibration gas cylinder (for either span or zero adjustment) and 

set the flow rate (see Table I).   

3. 

Connect the tubing to the regulator outlet and to the calibration adaptor (Figure II).  

 

 

REMOTE CALIBRATING THE S301M SENSOR 

When using the S301M in a network of sensors, connected to a 301C controller (with firmware 3.0 or 
higher), it is possible to perform a “remote calibration” using the 301C menu: 
Note: If the device is reading 0ppm, it is not necessary to adjust the zero. You can proceed to the span 
calibration. Remember to use Nitrogen whenever adjusting the unit’s zero. 
 

7. Network

Statistics
Calibration

 

 

1.  Using the navigation arrows, access the 301C’s menu 
2.  Scroll through the menu options and select option 7. Network 
3.  Scroll through the Network options screens to the last screen and select the Calibration option 

 

The Calibration screen contains four (4) lines of information: 
 

Calib 007

S301M

Status:

Normal

 

Set Zero

 

Set Span

246

PPM

 

 

Line 1:   Indicates the mode (Calib, meaning calibration), the (Modbus) address of the device to calibrate 

(007) and the type of device to calibrate (S301M) 

Line 2:   Indicates the status (Normal or In calib…) of the specified device* 
Line 3:   Displays the function to perform (Set Zero) 
Line 4:   Displays the function to perform (Set Span) and the span gas concentration value (246 ppm) 

 

1.  On the first line, scroll to the device address and press Enter 
2.  Scroll through the devices to display the desired device* and press Enter to select. 
3.  The second line displays the device’s status 
4.  Scroll to select the desired function, Set Zero to set the device's zero, and press Enter to select. 
 

Calib 007

S301M

Calibrate Zero?

 

Set Zero

 

Set Span

246

PPM

 

 

5.  Upon pressing Set Zero, the controller requests confirmation.   
6.  Press Enter to confirm or Esc to cancel.  If confirmed, the controller sets the S301M’s Zero.  This takes 

only a few moments and the display returns to the default calibration screen. 

 
* The display can also indicate a calibration error. If an error occurs, begin the process again from step 1. 

Содержание S301M

Страница 1: ... Scroll through the menu options and select option 7 Network 3 Scroll through the Network options screens to the last screen and select the Calibration option The Calibration screen contains four 4 lines of information Calib 007 S301M Status Normal Set Zero Set Span 246 PPM Line 1 Indicates the mode Calib meaning calibration the Modbus address of the device to calibrate 007 and the type of device ...

Страница 2: ...calib until the calibration is complete The device must be configured in the 301C s database in order to be included in the device addresses displayed on screen When selecting Set Span make sure that the device has been exposed to the appropriate calibration gas Table I Calibration Information Gas Scale Warm up Flow Rate C3H8 0 50 1 LEL 15 min 500 ml min CO 0 255 ppm 15 min 500 ml min NO2 0 10 2 p...

Страница 3: ...menu et sélectionner l option 7 Reseau 3 Dérouler les options de Reseau et sélectionner Etalonnage au dernier écran L écran d étalonnage contient quatre 4 lignes d information Calib 007 S301M Statut Normale Etal Zero Etal Plage 246 PPM Ligne 1 Indique le mode Calib l adresse de l unité pour l étalonnage 007 et le type d unité à étalonner S301M Ligne 2 Indique le statut Normale ou En Etalon de l un...

Страница 4: ...rre L écran affiche le statut de l unité En etalon jusqu à la terminaison de l étalonnage L unité doit être configurée dans la base de données du 301C pour figurer dans la liste d adresse d unité affichée à l écran Lors de la sélection d Etal Plage assurer que l unité a été exposée au gaz d étalonnage approprié Tableau II Information d étalonnage Gaz Plage Réchauffement Débit C3H8 0 50 1 LIE 15 mi...

Отзывы: