Clear Data:
Press the SET button to clear the
datalog. Press the MODE button to return to the
Datalog Set Menu.
Datalog Type:
Set the type of datalogging. Choices
are Continuous or Event-driven. Choose Continuous
to log data at a set interval at all times (except when
the unit is Programming Mode). Choose Event-
driven to log data only when the dose rate readings
increase or the unit goes into alarm.
Log Interval:
Set the time interval between datalog
points. You may set any time interval between 1
and 3600 seconds. Press the MODE button to step
through the digits from left to right. Use the SET
button to increment each digit (from 0 to 9).
Select Save to save changes or Quit to discard
changes.
Quit:
To exit Datalog Set, step through the options
using the MODE button until Quit is selected. Then
press the SET button.