Scale 

The Honeywell HomMed scale is available in regular (Wired) and Bluetooth (Wireless) versions. The scale has a capacity of 500 lb (227 kg) and 
collects weight using four load cells, one in each corner of the scale. The accuracy of the weight is +/- 0.5 lb (0.2kg). The Genesis DM supplies power 
and communication to the scale through the scale cord, or the scale can operate in a battery powered mode.  

NOTE: 

Bluetooth is a proprietary, open wireless method for exchanging data within relatively short distances using radio signals. For more 

information, see 

http://www.bluetooth.com

Gravity Adjustment:

 The scale is adjusted for gravity in the location to which HomMed is shipping it. Any relocation to a different gravity zone 

requires re-verification by HomMed. Any damage or recalibration may also require re-verification. The following table identifies distinguishing 
features of each scale. The Bluetooth scale does NOT have adjustable feet. 

 

 
 
 
 
 
 
 

 
 
The Honeywell HomMed scale is not a stand-alone device. The scale must be connected to a Honeywell HomMed 
monitor to operate. 

 
Step-by-Step Instructions covered in this guide are: 

 

 

Installing the Honeywell HomMed Bluetooth adapter 

 

Connecting one wireless scale 

 

Connecting when multiple wireless scales are in the area. 

 

Using the Scale during a Scheduled Reading 

 

Troubleshooting 

 

Feature 

 

Honeywell HomMed Wireless Scale 

Size 

Approximately 15.75"  by 14"   

Shape 

Curved 

Surface texture 

Round textured pattern 

Feet 

Fixed 

Scale cord  

4-pin MiniDIN Blue to monitor and  DB9 serial connector to scale  

Содержание HomMed

Страница 1: ...s DM Manual available while using the hardware This document is designed as a Quick Reference Guide only and it should be used in combination with the Genesis DM Manual and Telehealth Monitor Programming Troubleshooting Guide as it does not cover all aspects features or functions of the Genesis DM Telehealth Monitor This Quick Reference Guide is designed to help you perform the basic steps for add...

Страница 2: ...ive review of certain physiological functions The Genesis DM Monitor collects vital signs data including noninvasive blood pressure pulse rate and weight plus data from optional add on devices then can transmit the data to a central viewing station via a communication network The Genesis DM Monitor is intended for use with adult and pediatric patients over twelve years of age Contraindications The...

Страница 3: ...on to which HomMed is shipping it Any relocation to a different gravity zone requires re verification by HomMed Any damage or recalibration may also require re verification The following table identifies distinguishing features of each scale The Bluetooth scale does NOT have adjustable feet The Honeywell HomMed scale is not a stand alone device The scale must be connected to a Honeywell HomMed mon...

Страница 4: ...le is ready for use The configuration process would only need to be repeated if the monitor will be used with a different wireless scale or if the monitor will be used with the regular wired scale Installing the Honeywell HomMed Bluetooth adapter The wireless scale kit includes a Bluetooth Adapter and an Adapter Cable If needed securely connect the Adapter Cable to the Bluetooth Adapter Note they ...

Страница 5: ...ansmission and data flow between the wireless scale and the DM Monitor With the DM monitor power OFF plug the Adapter Cable into the Scale port at the back of the monitor or the wireless scale will not be sensed properly The adapter is equipped The adapter is equipped with light emitting diodes LEDs to indicate various conditions Blue Blinking Scanning for communicating with Bluetooth devices Soli...

Страница 6: ...ong with the request to press Yes to continue Power up the scale using either batteries or the AC power adapter supplied and place the scale in a position convenient for the patient Press YES to allow the monitor to detect the scale NOTE The scale should be less than 30 feet 10 meters from the monitor for signal acquisition The reason that the scale should be less than 30 feet 10 meters from the m...

Страница 7: ...nity The DM Monitor locates senses and pairs with a single scale automatically NOTE Establishing communications may take several minutes The Bluetooth adapter activity light emitting diodes will flash blue and red When the blue diode is flashing the Bluetooth devices in the vicinity are being scanned sensed and connected to the DM Monitor The blue LED is the Number One LED The red LED should be ON...

Страница 8: ... failed check to make sure that the scale has power and that the adapter is properly connected to the monitor and that the Bluetooth adapter red LED is ON Also verify that the scale is located close enough to the monitor and less than 10 meters or about 30 feet away After a failed attempt to detect a wireless scale the DM Monitor screen will ask if you desire a scan for available Bluetooth devices...

Страница 9: ...7EN 01 12 2010 9 When the scale is detected the monitor will display the Pairing Successful message Complete the Setup Wizard in the normal manner Scan for Available BT Devices YES or No Pairing Successful ...

Страница 10: ...hat you wish to pair with the monitor The serial number is located on the bottom of the scale You may write the Serial Number here for reference purposes ___________________________ After verifying the serial number power ON the DM Monitor When prompted enter the Setup Wizard by pressing YES The DM Monitor and Bluetooth Adapter will sense the presence of multiple devices Press YES to continue STAR...

Страница 11: ...teries or the scale AC power adapter Place the scale in a position convenient for the patient NOTE The scale should be less than 30 feet 10 meters from the monitor for signal acquisition Press YES to allow the monitor to detect the scale The monitor will display the Please Wait Scanning Devices message as it scans for suitable Bluetooth devices in the nearby vicinity Please Wait Scanning Devices P...

Страница 12: ...ear and prompt the user to select a scale to pair to the monitor Only one device or serial number is displayed on the monitor at a time Pressing the NO button for example goes down the list from 1 to 2 to 3 to 4 back to 1 Pressing the YES button goes up the list from 1 to the end of the list to 4 to 3 to 2 to 1 The RETEST button selects the currently displayed serial number IMPORTANT If a scale se...

Страница 13: ... is detected and paired with the monitor the monitor will display the Pairing Successful message After the Pairing Successful message appears complete the Setup Wizard in the normal manner The monitor will then be ready for use programmed with the wireless scale Select Scale To Pair 2 5094950020 Select Scale To Pair 3 5094950021 Pairing Successful ...

Страница 14: ...he user will be prompted to step on the scale as in normal wired operation If the monitor is configured for voice prompting the audible message will be Please step on the scale When the scale senses through a series of signal transactions that the person being weighed has stabilized on the scale then the weight will display The display of the weight may take a moment STEP ON SCALE o o o o o o o o ...

Страница 15: ...e message will be Please step off the scale This completes the installation and use of the Wireless Scale If you have any questions regarding the Wireless Scale please contact Honeywell HomMed Customer Service at 1 888 353 5404 STEP OFF SCALE o o o o o o o o o o 220 LBS ...

Страница 16: ...he monitor Solution Check the connection on the rear panel and retry E204 error code Bluetooth Adapter is not paired to the scale Solution Check the Bluetooth Adapter Make sure that the Bluetooth Adapter is powered ON and that it is properly connected to the Adapter Cable and that the Adapter Cable is properly connected to the DM Monitor at the Scale port on the rear panel of the DM Monitor Retry ...

Страница 17: ...e with four new AA batteries for best results NOTE The DM Monitor MUST BE RESTARTED after changing the batteries This is because the Bluetooth wireless scale is not configured to give a zero weight reading zeroed on power up The scale is only zeroed during monitor startup when the monitor does the scale self test This is unique to the Bluetooth wireless scale including when the AC adapter is being...

Отзывы: