MAINTENANCE, CLEANING AND REPAIR
Information for general use
Rev. M / 73023639
246
GasLab Q2
After several years of operation, additional maintenance and repair work may be required due to
wear and ageing.
Examples
of this include
battery replacement
(around 10 years)
and the
replacement of seals
and the
IR radiant heater
maintenance intervals at a glance
Furthermore, factory maintenance is required if defects are identified during the annual
maintenance work which cannot be rectified or if the housing has been damaged. Depending on
the situation, this work may require the complete dismantling of the device or its return to the
manufacturer. The required work may only be carried out by Honeywell service staff or by
personnel trained by Honeywell.
Exception:
Complete device replacement.
For example, the housing can only be repaired by the manufacturer since the dimensions of the
explosion-protection gaps are not public knowledge and are only specified in the
manufacturer’s
design documents. Please contact Technical Support if you require such services.
Explosion-protected devices which have not been repaired by the manufacturer must
not be reused until they have been inspected by an accredited expert and again comply
with the design documents.
L’équipement antidéflagration non réparé par le fabricant
ne doit pas être réutilisé jusqu’à ce qu’il ait été vérifié par un expert reconnu et qu’il soit à
nouveau conforme aux documents de construction.
Do not take any safety risks by using incorrect spare or replacement parts. Only genuine
Honeywell spare parts may be used.
No liability can be accepted for damage
caused by the use of non-genuine, non-approved spare and replacement parts.
10.5.
Device cleaning and documentation of the work
Ideally, you should use a damp cloth to clean the GasLab Q2. Generally, a mild cleaning solution
or a mild glass cleaner for domestic use may be used.
Do not use any equipment for cleaning which charges statically and may generate
sparks.
Do not pour or spray water or cleaning products directly into the breather element.
Do not use high pressure cleaners or similar equipment for cleaning.
Pour le nettoyage, n’utilisez que des outils qui ne permettront pas l’accumulation de
charge statique et ne créeront pas d’étincelles.
Il est interdit de verser ou de vaporiser de l’eau ou du détergent directement dans le
dispositif d’évent!
N’utilisez pas de nettoyeur haute pression ou d’équipement similaire pour le nettoyage!
Below you will find a form designed to help you document the completed work. You can copy this
form and add it to your documents.