Installation
SmartRadar FlexLine
Part No.: 4417.760_Rev06
Installation Guide
39
4.11.4.3 FII-RTD
4.11.4.3.1 RTD, 4-Wire Connection (Board ID = 2)
4.11.4.3.2 RTD, 3-Wire Connection (Board ID = 3)
Cable specifications
Type
Shielded. R
max
= 100
Ω
/ line
Maximum length
150 meter
Label
Connections diagram
Parameters
30. RTD1_S-
[Ex ia] IIB/IIC
U
o
= 8.61 V
I
o
= 58 mA
P
o
= 0.13 W
C
o
= 1.0
μ
F
L
o
= 10.5 mH
31. RTD1_com
32. RTD1_pos
33.
34. RTD2_S-
35. RTD2_com
36. RTD2_pos
37.
Label
Connections diagram
Parameters
30. RTD1_S-
[Ex ia] IIB/IIC
U
o
= 8.61 V
I
o
= 58 mA
P
o
= 0.13 W
C
o
= 1.0
μ
F
L
o
= 10.5 mH
31. RTD1_com
32. RTD1_pos
34. RTD2_S-
35. RTD2_com
36. RTD2_pos
RTD1
R
c1
RTD2
R
c2
RTD1
R
c1
RTD2
R
c2