background image

Honeywell EC-P2 Instruction and Installation Manual

15

4 Programming and Advanced Features

4.1 Alarm Setpoints, Functions

The  Honeywell  EC-P2  has  three  adjustable  alarm  setpoints. These  setpoints 
are stored on each SMART-CELL. The Caution setpoint is to warn of excessive 
negative zero drift. The WARNING and ALARM setpoints are to warn of potentially 
dangerous  target  gas  concentrations. The  following  procedure  will  change  the 
WARNING setpoint.

To Enter CAUT / WARN / ALRM Mode

Push power button to turn unit on. Wait for power up sequence to finish 

1. 

(30 seconds).
Push and hold the unnamed button first, then Push and hold the Menu 

2. 

button.  Hold  both  buttons  until  the  screen  goes  blank  (two  seconds). 
Release both buttons and the unit is now in Programming Mode.
Push the ALRM button and the unit is now in the CAUT / WARN / ALRM 

3. 

mode.

Warning Setpoint

Push WARN button.

4. 

Push S.P. button (setpoint).

5. 

Push  INC  button  (increase)  or  DEC  button  (decrease)  until  the  display 

6. 

matches the desired warning setpoint. Notice the minus sign above the 
arrow if the setpoint is below zero. 
Push SAVE button to store new warning setpoint.

7. 

Push DONE button to return to Alarms mode.

8. 

To Exit CAUT / WARN / ALRM Mode

Push DONE button twice slowly to get back into normal operation mode.

9. 

The Caution and Alarm setpoints are programmed in the same manner, except for 
step four above in which you would push either the CAUT or ALRM button. The 
function of each alarm can be selected to alarm on increasing gas concentration 
(UP arrow), alarm on decreasing gas concentration (DOWN arrow), or be disabled 
(off). The following procedure will set the Warning function to alarm on decreasing 
concentration (Useful for Oxygen Depletion warning).

To Enter CAUT / WARN / ALRM Mode

Push power button to turn unit on. Wait for power up sequence to finish 

1. 

(30 seconds).
Push and hold the unnamed button first, then Push and hold the Menu 

2. 

button.  Hold  both  buttons  until  the  screen  goes  blank  (two  seconds). 
Release both buttons and the unit is now in Programming Mode.
Push the ALRMS button and the unit is now in the CAUT / WARN / ALRM 

3. 

mode.

Содержание EC-P2

Страница 1: ...Honeywell EC P2 Instruction Manual CHARGE BATTERYWHEN NOT IN USE Gas Detector EC P2 ...

Страница 2: ......

Страница 3: ...en agreement with and for its customers In no event is Honeywell liable to anyone for any indirect special or consequential damages The information and specifications in this document are subject to change without notice Manning is a registered trademark of Honeywell International Inc Other brand or product names are trademarks of their respective owners Honeywell Analytics Inc 405 Barclay Bouleva...

Страница 4: ... carefully followed by all individuals who have or will have the responsibility for using or servicing this device Warranties made by Honeywell Analytics Inc with respect to this equipment will be voided if the equipment is not used and service in accordance with the instructions in this manual If in doubt about a procedure please contact Honeywell Analytics Inc before proceeding ...

Страница 5: ... Mode 13 Zero Calibration 13 Span Calibration 13 To Exit ZERO SPAN Mode 13 3 4 Troubleshooting 13 4 Programming and Advanced Features 15 4 1 Alarm Setpoints Functions 15 To Enter CAUT WARN ALRM Mode 15 Warning Setpoint 15 To Exit CAUT WARN ALRM Mode 15 To Enter CAUT WARN ALRM Mode 15 Warning Function 16 To Exit CAUT WARN ALRM Mode 16 4 2 Display Variables 16 To Enter AVG BLANK Mode 16 AVG Variable...

Страница 6: ... Working with the EC P2 program 19 4 5 Analog Output 19 Changing the full scale range on EC P2 20 To Enter RANGE Mode 20 Adjusting the RANGE 20 To Exit RANGE Mode 20 Appendix A 21 Available SMART CELLs 21 Appendix B 22 SMART CELL Cross Sensitivity Table Interference Gases 22 Limited Warranty 24 ...

Страница 7: ... two different batteries to insure that you ll never be without power when you need it The primary power source is a D size alkaline battery housed in the handle It will run the unit approximately 75 hours When this battery is removed or dead the unit will switch to its standby power source an internal rechargeable NiCad battery which will run the unit approximately six hours When not in use the u...

Страница 8: ... Optional Accessories Analog output cable Sensor Keeper for storage of additional SMART CELLs Sensor Type Target gas selective electrochemical SMART CELL Accuracy Sensor dependant but generally 5 of reading Repeatability 1 full scale Operating Ambient Temperature Range 0 to 120 F Storage Temperature Recommend storing on the charger in an air conditioned office Weight 3 lbs shipping weight 1 8 lbs ...

Страница 9: ...can be performed with the unit powered on or off see Figure 3 Unscrew the two stainless steel thumbscrews and remove the black inlet 1 outlet manifold lid from the back of the unit To remove the existing SMART CELL grip the sides of the cell and pull 2 it straight out Install the desired SMART CELL by gently inserting it into the socket Notice 3 that the SMART CELL is keyed so it cannot be inserte...

Страница 10: ... show an increasing concentration If not do not use it for field measurements 2 4 Sample Inlet Port The Honeywell EC P2 is provided with a quick disconnect flexible extension wand It will work with or without the wand When using the extension wand the internal walls must be kept dry Water on the walls of the tubing can potentially absorb the target gas To dry allow the unit to pump dry ambient air...

Страница 11: ... Alarm Indication Code will remain on the left side of the screen until the alarm condition clears Alarm setpoints are stored on the individual SMARTCELL and can be changed by the user as described in Section 4 Note An audible alarm single beep every 60 seconds indicates that either the Alkaline D cell battery is low or the internal rechargeable NiCad battery is low 2 8 Display Resolution The Hone...

Страница 12: ...at the sensor will not respond For this reason it is absolutely essential that these units be exercised with a gas sample on a regular basis Typical sensor life under normal use is two years or more When the SMART CELL will no longer calibrate properly simply purchase a new SMART CELL and replace 3 3 Calibration Calibration is recommended every six months in normal use Each SMART CELL stores its c...

Страница 13: ... span gas for two minutes 8 Push INC button increase or DEC button decrease until the display 9 matches the span gas concentration Push SAVE button to store new span calibration 10 To Exit ZERO SPAN Mode Push DONE button twice slowly to get back into normal operation mode 11 3 4 Troubleshooting The detector does not start Check to see that a fresh battery is installed Both the primary and standby ...

Страница 14: ...r to clear Leave the unit running in a clean environment for a few hours to clear the sensor If it doesn t clear after six hours the sensor may have been damaged Contact Honeywell Analytics for a replacement SMART CELL Unit is in alarm with gas reading of 0 ppm Check the alarm setpoints and alarm functions as described on page 12 Verify that alarm is not programmed as a downscale alarm and that th...

Страница 15: ...until the display 6 matches the desired warning setpoint Notice the minus sign above the arrow if the setpoint is below zero Push SAVE button to store new warning setpoint 7 Push DONE button to return to Alarms mode 8 To Exit CAUT WARN ALRM Mode Push DONE button twice slowly to get back into normal operation mode 9 The Caution and Alarm setpoints are programmed in the same manner except for step f...

Страница 16: ... Display Variables The Honeywell EC P2 has two adjustable display variables These variables control signal conditioning of the LCD display Variables are stored on each SMART CELL The AVG Average variable determines the number of one second samples to average to calculate the displayed value The lower the averaging value the faster the display will change and the more wiggle will be displayed The B...

Страница 17: ...med measuring time Clear the gas concentration to below a user programmed recovery value before allowing another cycle to start Unit should be moved to fresh air during the clear time Taking a sample To take a sample locate the detector at the desired sample location and push the SMPL sample button on the normal operation screen The unit will count down the Sample time followed by the Measure time...

Страница 18: ...upplied with a CD containing the Honeywell EC P2 program It will run on Windows 98 Windows 2000 or Windows XP To install the program place the disk in your CD drive The CD should automatically begin the installation process Follow the prompts as they appear on your screen Accepting the default directory locations is recommended for most installations If the CD does not automatically start click on...

Страница 19: ...AMS then Honeywell EC P2 and then EC P2 As soon as you start the EC P2 program your computer will try to establish a connection to the Honeywell EC P2 If the Honeywell EC P2 is currently logging data the EC P2 program will automatically stop the logging function The program contains two tabs one marked Download and one marked Configure Review the settings on the Configure tab making changes as nec...

Страница 20: ... unit on Wait for power up sequence to finish 1 30 seconds Push and hold the unnamed button second from left first then Push 2 and hold the MENU button Hold both buttons until the screen goes blank two seconds Release both buttons and the unit is now in Programming Mode Push the MORE button 3 Push the DISP display button 4 Push the MORE button and the unit is now in the RANGE mode 5 Adjusting the ...

Страница 21: ...ppm 0 1 ppm 10944 Fluorine F2 0 1 ppm 0 5 ppm 0 01 ppm 10174 Hydride gases 0 500 ppb 0 2000 ppb 1 ppb 10945 Hydrogen H2 0 500 ppm 0 2000 ppm 1 ppm 10176 Hydrogen Peroxide H2 O2 0 10 ppm 0 100 ppm 1 ppm 10178 Hydrogen Sulfide H2 S 0 10 ppm 0 200 ppm 0 1 ppm 10179 Formaldehyde CH2 O 0 20 ppm 0 200 ppm 0 1 ppm 10175 Hydrogen Chloride HCl 0 10 ppm 0 200 ppm 0 1 ppm 10177 Hydrogen Cyanide HCN 0 10 ppm ...

Страница 22: ...1 1 0 4 0 4 0 01 0 5 1 Formaldehyde None None None None 0 03 0 1 0 1 None 0 05 0 01 None 0 02 1 2 None None 0 1 Alcohol None None None None 0 03 0 1 0 1 None 0 05 0 01 None 0 02 None None None 0 1 Flourine 1 0 1 0 1 0 0 7 0 1 None None None None None 0 5 0 1 None None None None Hydrochloric acid None None None None None None None None None None 0 3 None 0 2 0 2 None 0 1 Hydrogen peroxide None None...

Страница 23: ...sensor 2 will react the specified amount for a given concentration For example on the first line of the table above if the NO2 sensor is exposed to 1 ppm Bromine Br2 the sensor will read 0 5 ppm This interference is in addition to the target gas present For example if an NH3 sensor is exposed to 1 ppm NH3 and 1 ppm Br2 at the same time the reading will be 0 9 ppm ...

Страница 24: ...warranty This limited warranty does not cover C consumable items such as batteries or items subject to wear or periodic adjustment including lamps fuses valves vanes sensor elements cartridges or filter elements Warranty Limitation and 3 Exclusion Honeywell Analytics will have no further obligation under this limited warranty All warranty obligations of Honeywell Analytics are extinguishable if th...

Страница 25: ...Honeywell EC P2 Instruction and Installation Manual 25 ...

Страница 26: ......

Страница 27: ......

Страница 28: ...208 detectgas honeywell com Technical Services ha global service honeywell com www honeywell com Please Note While every effort has been made to ensure accuracy in this publication no responsibility can be accepted for errors or omissions Data may change as well as legislation and you are strongly advised to obtain copies of the most recently issued regulations standards and guidelines This public...

Отзывы: