![Honeywell DR4300 Series Скачать руководство пользователя страница 189](http://html.mh-extra.com/html/honeywell/dr4300-series/dr4300-series_product-manual_151814189.webp)
Troubleshooting and Pen Alignment
of Recorder with Display
4/00
DR4300 Circular Chart Recorder Product Manual
177
Table 8-4 Error Messages
Lower Display
Indication
Type of Test
Reason For Failure
How to Correct the Problem
CALERR
power-up
background
The working calibration constants
in the recorder are in error or
corrupted in memory.
1. Use the procedure in Subsection 4.5.2.5
to switch between the set of calibration
constants being used (factory or field) to
the other type. Exit configuration mode.
Next go back into configuration mode
and change back to the original type of
constants used (factory or field). This will
overwrite the corrupted values in
memory.
2. Change the setting of the “IN TYP”
parameter and exit configuration mode.
Go back into configuration mode and
change the “IN TYP” setting back to the
correct type.
CFGERR
power-up
background
Configuration data is in error.
1. Check all configuration parameter values
for accuracy. (See Section 4 –
Configuration, Startup, and Operation of
Recorder with Display.)
2. Make a change to configuration, then exit
configuration, so the recorder will
recalculate the checksum error.
EEFAIL
background
Unable to write to nonvolatile
memory. Anytime you change a
parameter and it is not accepted,
you will see EE FAIL.
1. Try to change something in configuration.
2. Cycle the power.
3. If problem persists, replace printed circuit
assembly for the channel.
FAILSF
power-up
background
This error message is displayed
whenever the recorder is using
the value of the control group
“FAILSF” parameter as an output.
For this to occur, the recorder
must have control enabled
(control group parameter
“CONTRL” = “ENAB”) when one
or more of the following occur:
• RAM test fails
• Configuration test fails
• Calibration test fails
• Input parameter
“BRNOUT” = “NONE” and the
input fails.
1. Check the operation of the input sensor.
2. Run through STATUS check to
determine the reason for the failure.
• Press the SET UP key until
STATUS appears in the lower display.
• Press the FUNC key to see what
tests pass or fail, then run through the
STATUS codes a second time to see
if the error cleared.
• Correct according to the
recommendations given in this table.
Содержание DR4300 Series
Страница 1: ...Sensing and Control DR4300 Circular Chart Recorder Product Manual 44 01 25 14E 4 00 ...
Страница 12: ...DR4300 Circular Chart Recorder DR4300 Circular Chart Recorder Product Manual 4 00 xii ...
Страница 82: ...DR4300 Circular Chart Recorder 70 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 152: ...DR4300 Circular Chart Recorder 140 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 168: ...DR4300 Circular Chart Recorder 156 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 202: ...DR4300 Circular Chart Recorder 190 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 210: ...DR4300 Circular Chart Recorder 198 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 214: ...DR4300 Circular Chart Recorder 202 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 224: ...DR4300 Circular Chart Recorder 212 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 230: ...DR4300 Circular Chart Recorder 218 DR4300 Circular Chart Recorder Product Manual 4 00 20766 20766 20766 ...
Страница 238: ...DR4300 Circular Chart Recorder 226 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 244: ...DR4300 Circular Chart Recorder 232 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 245: ...Appendix E Foreign Language Safety Instructions 4 00 DR4300 Circular Chart Recorder Product Manual 233 ...
Страница 246: ...DR4300 Circular Chart Recorder 234 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 247: ...Appendix E Foreign Language Safety Instructions 4 00 DR4300 Circular Chart Recorder Product Manual 235 ...
Страница 248: ...DR4300 Circular Chart Recorder 236 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 249: ...Appendix E Foreign Language Safety Instructions 4 00 DR4300 Circular Chart Recorder Product Manual 237 ...
Страница 250: ...DR4300 Circular Chart Recorder 238 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 251: ...Appendix E Foreign Language Safety Instructions 4 00 DR4300 Circular Chart Recorder Product Manual 239 ...
Страница 252: ...DR4300 Circular Chart Recorder 240 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 253: ...Appendix E Foreign Language Safety Instructions 4 00 DR4300 Circular Chart Recorder Product Manual 241 ...
Страница 254: ...DR4300 Circular Chart Recorder 242 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 258: ...DR4300 Circular Chart Recorder 246 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 264: ...252 DR4300 Circular Chart Recorder Product Manual 4 00 ...
Страница 265: ......