Honeywell BW Solo User Manual
12
3.2.
Status Indicator Icons
On-screen icons are used to indicate battery, bump, calibration and sensor status, as well as peak
reading. The table below details how and when each icon is displayed. The triangular red LED
located to the right of the screen illuminates to accompany these (it is active when the non-
compliance feature is on).
Icon
Explanation
The battery icon is not shown on the display unless the remaining power is
one day or less.
•
When the battery’s remaining life falls below 24 hours, the battery icon is
shown and there is a flash and beep every 60 minutes.
•
When the battery’s remaining life falls below 8 hours, the icon blinks on
and off every second, and there are two beeps and a flash every 30
seconds.
When less than 12 hours remains before a bump test should be performed,
this icon is shown and the buzzer sounds once every hour. When the user
has also set the non-compliance option for the bump test reminder, the
instrument beeps once every 10/30/60 seconds (the interval is set by the
user).
The icon blinks once per second if there is a failed or skipped bump test, and
the instrument beeps three times every 60 minutes.
A bump test is required (and indicated by this icon) if:
•
The defined period between bump tests has been exceeded (bump test
overdue).
•
The sensor has failed a previous bump test.
Failed bump test notification is shown on the screen if a bump test is failed. In
the information menu, it shows that the bump test must be performed “NOW.”
Note:
the sensor should be challenged on a regular basis (before each day’s
use) to ensure proper sensor performance.
This icon is shown when there is less than 12 hours remaining before a
calibration should be performed. The instrument beeps once every 60 minutes.
The icon blinks once per second if there is a failed or skipped calibration, and
the instrument beeps three times every hour if the non-compliance option is
on. When the user has also set the non-compliance option for the calibration
reminder, the instrument beeps once every 10/30/60 seconds (the interval is
set by the user).
Calibration is required (and indicated by this icon) if:
•
The defined period between calibrations has been exceeded.
•
The sensor has failed a previous calibration.
When a sensor fails, has been removed, or has expired, this icon is shown on
the display, and there are is one beep per second. Stepping through the status
screens, you see the fault details.
The Peak Reading icon is displayed when a peak gas reading has been
detected in the past 24 hours. The icon is otherwise hidden.