Calibration
Use this menu to perform a bump test or zero or span calibration for one or more sensors, and
change the gas concentration value used in bump tests and span calibration, as well as choose
which sensors will be calibrated at the same time.
Fresh Air
This procedure determines the zero point for all the sensors that require a zero calibration. For
the oxygen sensor, Fresh Air calibration sets the point equal to the concentration of oxygen in
ambient air (approximately 20.9% volume).
Note: Fresh air calibration is performed on all enabled gas sensors at the same time.
To perform Fresh Air calibration on multiple sensors:
1.
Connect the instrument’s inlet to a source of dry, clean ambient air.
2.
At the Calibration Menu, select “Fresh Air.” Press [Y/+] once to enter the fresh air
calibration sub-menu.
3.
Start the flow of dry zero air, if used.
4.
Press [Y/+] to start fresh air calibration.
5.
A countdown screen appears. You can abort the calibration at any time during the
countdown by pressing [N/-].
Note: Dotted line indicates automatic progression.
6.
If the calibration is not aborted, the display shows the sensor names and tells you
whether the fresh air
Multi Sensor Span
Depending on the configuration of your AreaRAE Plus/Pro and the span gas you have, you can
perform a Span calibration simultaneously on multiple sensors.
The selected sensors are shown on the screen, along with the concentration settings for their
Span gas. With calibration gas connected to the instrument, start a multi span by applying gas to
the instrument. Calibration should start after a few seconds. If not, press [Y/+]. If you do not want
to perform a multi span, press [MODE].
Note: You can abort a multi span by pressing [N/-] once testing has started.
When the Multi Span is done, a screen is shown, with the sensor names and either “Pass” or
“Fail” shown next to them.
AreaRAE
54
User Manual
Содержание AreaRAE Plus
Страница 1: ...AreaRAE WIRELESS MULTI GAS MULTI THREAT DETECTOR ...
Страница 2: ......
Страница 17: ...AreaRAE 16 User Manual ...
Страница 22: ...Temperature Relative humidity AreaRAE 21 User Manual ...
Страница 34: ...The instrument should wake up and begin monitoring immediately AreaRAE 33 User Manual ...
Страница 106: ...5 Once the cap is removed set it aside 6 Now lift the sensor electrode panel from the module AreaRAE 105 User Manual ...
Страница 145: ...Rev C September 2021 User Manual P N W01 4001 000 ...