Table 2
Guidance and manufacturer’s declaration – electromagnetic emission
Immunity test
IEC 60601 test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV, air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV, air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0%UT; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°,225°,270° and 315°. 0%UT ; 1 cycle
and 70%UT ; 25/30 cycles. Single phase: at
0°. 0% UT ; 300 cycle
0%UT; 0.5 cycle. At 0°, 45°, 90°, 135°,
180°,225°,270° and 315°. 0%UT ; 1 cycle
and 70%UT ; 25/30 cycles. Single phase: at
0°. 0% UT ; 300 cycle
Power frequency
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conducted RF
IEC61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
17
I
GB
16
I
GB