HOMEDICS
meˉ
18
GB
Troubleshooting
Indicator
What does it mean
How to react
Control Panel
No lights are on
Power cable is not connected Connect power cable
All elōs energy level lights are
blinking
No elōs energy level was
selected .
Select an elōs level. The
applicator fan will start blowing
indicating that the cartridge is
active .
elōs activation button is
blinking
elōs energy level was
selected. elōs activating
button flashes to indicate that
it needs to be pressed .
Press the cartridge activation
button . The applicator fan will
start blowing indicating that the
elōs cartridge is active.
Power light is blinking orange
An error has been detected
Reset:
1 . Disconnect power cord .
2 . Wait until all lights are off (up
to 2 minutes).
3 . Reconnect the power cord
and turn system on .
4 . Move to a cooler area with a
room temperature of less than
86 degrees Fahrenheit (30
degrees Celsius) and restart
the system .
No light flashes from cartridge
treatment window .
1 . RF skin contact bars not
in contact with the skin in
treatment area .
2 . Cartridge activation button
is n ot pressed .
In order for the system to flash
pulses, make sure all 3 conditions
are met: RF skin contact bars
in contact with skin, cartridge
activation button on, elōs level
selected .