38
VER: A001 202203
Appendix
Test specifications for enclosure port immunity to RF wireless
communications equipment.
Test frequency
(MHz)
Modulation
IMMUNITY
TEST LEVEL
(V/m)
385
Pulse modulation 18 Hz
a)
27
450
FM
±
5 kHz deviation 1kHz sine
b)
28
710
Pulse modulation 217 Hz
a)
9
745
780
810
Pulse modulation 18 Hz
a)
28
870
930
1720
Pulse modulation 217 Hz
a)
28
1845
1970
2450
Pulse modulation 217 Hz
a)
28
5240
Pulse modulation 217 Hz
a)
9
5500
5785
NOTE:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m, The 1 m test
distance is permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case.