table 3 Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUiPMEnt and ME sYstEMs that are not liFE-sUPPortinG
Guidance and manufacture’s declaration – electromagnetic immunity
the BPA-2000-EU is intended for use in the electromagnetic environment specified below. the customer or
the user should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
conducted rF
iEc 61000-4-6
radiated rF
iEc 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m 80 MHz
to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile rF communications
equipment should be used no closer to any
part of the Ys-6100, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.167
√—
P
d = 1.167
√—
P
80 MHz to 800 MHz
d = 2.333
√—
P
800 MHz to 2.5 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the recommended
separation distance in metres (m).
Field strengths from fixed rF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
interference may occur in the vicinity of
equipment marked with the following
symbol:
notE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
notE 2 these guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and tV broadcast cannot be predicted
theoretically with accuracy.
a
to assess the electromagnetic environment due to fixed rF transmitters, an electromagnetic site survey
should be considered. if the measured field strength in the location in which the ElE007839V1 is used
exceeds the applicable rF compliance level above, the ElE007839V1 should be observed to verify normal
operation. if abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the ElE007839V1.
b over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.