Discovery QDR Series Technical Manual
3-16
Figure 3-18. Unit in Vertical Position
3. Remove the auxiliary caster bars.
4. Move the unit and everything that shipped with it to the destination room.
3.2.9 Set Up the Unit
Follow the procedure below to set up the unit:
1. Carefully tilt the unit down.
2. Remove the caster bars and cross brace. Position the unit in its final location in
the room.
3. Level the unit.
4. Remove the 2 carriage-to-base shipping brackets (Figure 3-19).
5. Discovery A, W & Wi - Remove the table end stop (Figure 3-10).
6. Remove the table locking bracket located on the left rear side of the scanner
(Figure 3-9).
WARNING:
The unit will feel heavier when putting it down than it felt when tipping it up. This is
because the pivot point is different. For safety purposes it is recommended that two
people are used to tilt the unit down.