Affirm User Guide
Chapter 4:
How to Use the System
MAN-03209 Revision 005
Page 35
4.3.2 C-Arm Stereo Modes
Acquire the stereo images in either the Auto C-Arm Stereo Mode or the Manual C-Arm
Stereo Mode. Make the C-Arm Stereo Mode selection at the Acquisition Workstation
(refer to
How to Select the C-Arm Stereo Mode for Image Acquisition
on page 36).
C-Arm Rotation in the C-Arm Stereo Modes
The mechanics of C-arm rotation for stereotactic image acquisition is different in Auto and
Manual C-Arm Stereo Modes.
Note
C-arm vertical movement and rotation are disabled when
compression force is applied. A service engineer can
configure the lockout force from 22 Newtons (5 pounds) to
45 Newtons (10 pounds).
In Auto C-Arm Stereo
Mode
In the Auto C-Arm Stereo Mode, the Target Guidance screen does not show the
C-Arm
Stereo Mode
button. In the Auto C-Arm Stereo mode, the system automatically lets the
C-arm move separately from the Tube Arm to acquire the stereo images. When the
X-ray
button is pressed to start stereo image acquisition, the Tube Arm automatically moves to
the first 15° angle, acquires the image, rotates to the opposite 15° angle and acquires that
image.
Note
The option to position the Tube Arm for the first stereo
image is in the System Status menu (refer to
C-Arm Stereo Mode for Image Acquisition
on page 36). The
Tube Arm automatically rotates to the selected 15° angle.
This option operates in both C-Arm Stereo modes.
Содержание Affirm
Страница 1: ......
Страница 2: ......
Страница 3: ...User Guide For Software Version 1 7 Part Number MAN 03209 Revision 005 January 2015...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 32: ......
Страница 54: ......
Страница 58: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 69: ......
Страница 70: ......