In Manual C-Arm Stereo Mode
In Manual C-Arm Stereo Mode, the
C-Arm Stereo Mode
button shows on the Target
Guidance screen. The default is Stereo Mode. See the table that follows for a description of
the button function.
Note
The C-Arm Stereo Mode button shows when the system is
set for Manual C-Arm Stereo Mode. In Auto C-Arm Stereo
Mode, this button does not show.
Table 4:
The C-Arm Stereo Mode Button
Icon
Description of Function
Figure 24:
Stereo Mode
•
The Tube Arm rotates while the C-Arm remains in
position.
•
Select Stereo Mode to acquire stereo images (when
the system is set for Manual C-Arm Stereo Mode).
Figure 25:
C-Arm Mode
•
The C-arm and Tube Arm rotate together.
•
Select C-Arm Mode to rotate the C-arm and Tube
Arm together to a new patient positioning angle.
1.
After you set the patient positioning angle (with the C-Arm Mode engaged), press the
C-Arm Stereo Mode
button on the Target Guidance screen to put the C-arm into
Stereo Mode.
2.
Rotate the Tube Arm to the first 15° angle.
3.
Press the
X-ray
button and acquire the first 15° image.
4.
Rotate the Tube Arm to the opposite 15° angle.
5.
Press the
X-ray
button and acquire the opposite 15° image.
Note
Keep the
X-ray
button pressed until the exposure stops. An
audible beep indicates that image acquisition has finished.
Содержание Affirm
Страница 1: ......
Страница 2: ......
Страница 3: ...User Guide For Software Version 1 7 Part Number MAN 03209 Revision 005 January 2015...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 32: ......
Страница 54: ......
Страница 58: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 69: ......
Страница 70: ......