Acessa

®

 ProVu System User’s Guide

 

PL-01-0040 

 

 

CO 20-131  Revision: C   

38

 

 

Condition 

Visual Alert 

Audible 

Alert 

Solution 

Pad contact resistance high during 

ablation 

"RF output ended, high pad 

contact resistance" 

Yes 

A high pad contact resistance was 

detected. 

1. Check the pads to ensure that they 

are in good contact with the patient. 

2. Replace the pads (if necessary. 

Unrecoverable error (e.g. failed SBC to 

RF communications) 

"Instrument failure.  Please 

power down the machine" 

if 

possible 

1. Press and hold the Standby 

pushbutton for about 5 seconds and 

then release it. 

2. Then turn off the rear power switch.   

3. Wait at least 10 seconds and then 

restart the Console. 

4. If it happens again, call Acessa 

Health with failure number. 

Instrument encounters a fatal error 

condition. 

An error number and error text 

will be displayed 

if 

possible 

1. Press and hold the Standby 

pushbutton for about 5 seconds and 

then release it. 

2. Then turn off the rear power switch.   

3. Wait at least 10 seconds and then 

restart the Console. 

4. If it happens again, call Acessa 

Health with failure number. 

Guidance system failure 

The software has encountered an error, 

and the user is given a code to report to 

Tech Support 

“Guidance Failure <varies>. 

Contact Technical Support” 

Yes 

Call Acessa Health with failure number. 

 

 

 

Содержание Acessa ProVu

Страница 1: ...Acessa ProVu System User s Guide PL 01 0040 CO 20 131 Revision C 1 USER S GUIDE Release Date 2 5 2021...

Страница 2: ...eration of System During Coag 19 10 13 After the Procedure 19 10 14 Shutting Down the System 19 11 Special Considerations General Ablation Procedures 20 11 1 Highly Vascularized Tissue 20 11 2 One or...

Страница 3: ...ling 31 17 Care and Maintenance 31 17 1 Software Upgrades and Installation 31 17 2 Maintenance 31 17 3 Cleaning and Disinfecting the Console 31 17 4 Calibration Verification 31 18 Specifications 31 18...

Страница 4: ...he ablation throughout the procedure The Acessa ProVu System must be used under laparoscopic ultrasound guidance The basic function of ultrasound is to acquire ultrasound echo data and to display the...

Страница 5: ...Acessa ProVu System User s Guide PL 01 0040 CO 20 131 Revision C 4...

Страница 6: ...the patient may result in burns Due to concerns about the carcinogenic and infectious potential of electrosurgical byproducts such as tissue smoke plume and aerosols protective eyewear filtration mask...

Страница 7: ...ses within body cavities or originate in surgical drapes tracheal tubes or other materials ASPIRATE fluid from the area before activating the instrument Conductive fluids e g blood or saline in direct...

Страница 8: ...f monopolar electrosurgery particularly in the prevention of burns Follow directions and recommended practices for the preparation placement surveillance removal and use of the Pads Use with the syste...

Страница 9: ...communications equipment can affect the Acessa ProVu Systemfunctionality Do not expose the Handpiece or Transducer with Sensorto a high magnetic field such as a Magnetic Resonance Imaging MRI scanner...

Страница 10: ...use Clean and disinfect the Pad Cable in accordance with the recommended process in the Instructions for Use accompanying the cable Ensure that the Pad cable interconnections are clean and dry prior t...

Страница 11: ...r Connector Accepts the connector from either the Table Top Field Generator or Planar Field Generator K Handpiece Connector Accepts either end of the Handpiece Cable L Return Pad Connector Accepts the...

Страница 12: ...ed R is connected to the wye marked R and similarly L to L for the other pad Affix the Return Pads to the patient per the Pad IFU PL 01 0015 Consult Pads Instruction for Use for prep of patient and pr...

Страница 13: ...ssories can be connected before or after the systemhas been powered ON except TTFG cable and power cord which must be connected prior to system power ON 9 4 Energizing The System Place the power switc...

Страница 14: ...atient s feet surgeon is on the patient s right The left side shows Guidance and the right side shows the ultrasound The screen layout is most efficient if the ultrasound side is closest to the laparo...

Страница 15: ...evision C 14 10 3 User Interface Upper Left Deployment Ablation Coag Center Top Expected Ablation Size Upper Right Ultrasound Settings Center Middle Proximity Meter Lower Left Menu Selections Lower Ce...

Страница 16: ...tions are 3cm 4cm 5cm 6cm 7cm 8cm and 9cm Frequency The Frequency penetration resolution options are 5MHz 6MHz 9MHz and 12MHz Gain The Gain brightness options are 0 100 in 1 increments Focus The Focus...

Страница 17: ...al to the level setting The user slowly withdraws the Handpiece while visually monitoring the coag laparoscopically adjusting level setting as necessary to achieve optimal results Turn off RF power wh...

Страница 18: ...ot be taken using them If the Handpiece is inserted out of plane do not use the meter reference the 2D display Handpiece ultrasound Target Zone The software provides a prediction of the Handpiece s pa...

Страница 19: ...advance the Handpiece until the tip is at the serosa of the uterus viewed on the laparoscopic monitor After the tip penetrates the serosa view image on the ultrasound monitor Corrections to the path...

Страница 20: ...e last setting chosen The Handpiece tip should be allowed to cool for 60 seconds after the ablation has stopped prior to removing it from the target tissue When ready to Coag turn the RF energy on by...

Страница 21: ...emperature for stated target time for the last deployment stage to allow for complete tissue destruction If a fibroid is densely concentrated with calcium to the degree that there is abnormal resistan...

Страница 22: ...tatus was gravida 2 8 2 0 and para 2 1 1 6 At the time of screening 81 of the subjects reported having regular menstrual cycles and 77 reported having heavy to very heavy bleeding volume Nearly 85 of...

Страница 23: ...nths Subjects were required to have a normal coagulation profile a normal Pap smear and pass a preoperative health examination American Society of Anesthesiologists ASA physical status I III Subjects...

Страница 24: ...In cases where deployment of the electrode array was performed the array was deployed so that the electrode tips were at least 1 cm away from the fibroid margin in all three planes Laparoscopic ultra...

Страница 25: ...at 12 months post treatment Subject Disposition at 12 Months Post Treatment n Total Treated 137 100 Evaluable for Menstrual Blood Loss Using Alkaline Hematin 124 90 5 Withdrew prior to 12 months 2 Lo...

Страница 26: ...8 Efficacy Menstrual Blood Flow as Assessed by Alkaline Hematin AH In the evaluable subjects n 124 the mean reduction in menstrual blood flow at 12 months post treatment was 103 6 ml One hundred four...

Страница 27: ...oncern Activities Energy Mood Control Self Consciousness and Sexual Function Scores at 3 6 12 24 and 36 Months Post Treatment 16 25 38 44 60 50 40 30 20 10 0 Month 3 Month 12 Change From Baseline Uter...

Страница 28: ...ecame pregnant in the second year of follow up One pregnancy resulted in a spontaneous abortion in the first trimester one progressed normally resulting in a vaginal delivery and two subjects had a C...

Страница 29: ...Sites 14 sites in the United States and Canada enrolled participants in the study 14 4 Total Number of Enrolled Subjects and Follow up Rate A total of 110 participants were enrolled Five 5 subjects w...

Страница 30: ...ts who participated in that follow up visit For the acute serious complication rate the null hypothesis was that the acute serious complication rate for the Acessa procedure is no different than the o...

Страница 31: ...n mild intra operative bleeding transient urinary retention or urinary tract infection adhesion formation post procedural discomfort cramping pelvic pain and transient amenorrhea The risk of skin burn...

Страница 32: ...ld be given reasonable care and be kept clean and sanitary To clean the Console Foot Pedal and Field Generators wipe down the devices with a 70 isopropyl alcohol wipe Clean surfaces until all visible...

Страница 33: ...ad Port Transducer Port Transducer Sensor Port Field Generator Port USB Port Video Port HDMI Power Entry Module with internal fuses Protection Class I Type BF Applied Part Rated for ordinary Continuou...

Страница 34: ...5 0 5kV 1 kV for line to line 0 5kV 1 kV 2kV for line to ground 0 5kV 1kV line to line 0 5kV 1kV 2kV line to ground Mains power quality should be that of a typical commercial or hospital environment...

Страница 35: ...e level above the system should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the system b Over the...

Страница 36: ...ation lower levels may be more effective Watts of Power vs Rload Ablate Modes Note The unit is limited at low load impedances by maximum available current and is limited at high load impedances by max...

Страница 37: ...d Errors The system detects and displays fault and error conditions A fault is a correctible condition that produces an Alert screen while an error is irreversible and requires system reboot to clear...

Страница 38: ...unplugged Yes 1 Press Acknowledge Check Button 2 Check Pad or cable connection 3 If connections are good replace Cable 4 If Cable was good replace Pad Fewer than 4 valid functioning Handpiece TCs Han...

Страница 39: ...ds and then release it 2 Then turn off the rear power switch 3 Wait at least 10 seconds and then restart the Console 4 If it happens again call Acessa Health with failure number Instrument encounters...

Страница 40: ...Not suitable for use with flammable anaesthetics IEC 60601 1 2012 Table D 2 Symbol 10 Follow Instructions for Use Refer to instruction manual booklet IEC 60417 5036 Warning Electricity Warning danger...

Страница 41: ...needs to be protected from moisture IP2X IEC 60529 IP2X Protection against approach by fingers ISO 60601 2 1 2009 IEC 1193 06 Figure 201 102 2009 HF Isolated Patient Circuit High Frequency Isolate Pat...

Страница 42: ...ring or unauthorized servicing may void any manufacturer s warranties LIMITATIONS and OBLIGATIONS THE FOREGOING WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES AND IMPLIED WARRANT...

Отзывы: