38
Condition
Visual Alert
Audible
Alert
Solution
Pad contact resistance high during
ablation
"RF output ended, high pad
contact resistance"
Yes
A high pad contact resistance was
detected.
1. Check the pads to ensure that they
are in good contact with the patient.
2. Replace the pads (if necessary.
Unrecoverable error (e.g. failed SBC to
RF communications)
"Instrument failure. Please
power down the machine"
if
possible
1. Press and hold the Standby
pushbutton for about 5 seconds and
then release it.
2. Then turn off the rear power switch.
3. Wait at least 10 seconds and then
restart the Console.
4. If it happens again, call Acessa
Health with failure number.
Instrument encounters a fatal error
condition.
An error number and error text
will be displayed
if
possible
1. Press and hold the Standby
pushbutton for about 5 seconds and
then release it.
2. Then turn off the rear power switch.
3. Wait at least 10 seconds and then
restart the Console.
4. If it happens again, call Acessa
Health with failure number.
Guidance system failure
The software has encountered an error,
and the user is given a code to report to
Tech Support
“Guidance Failure <varies>.
Contact Technical Support”
Yes
Call Acessa Health with failure number.