User Manual
25
Version 1.0
Fault
Possible Reason
Remedy
Noise is Abnormal
The oil level of the pump is
too low
Add the oil to the level as advised
The pump inner parts are
damaged
Disassemble and replace parts
Poor ventilation
Improve ventilation clearance around the pump
The fan is damaged
Change the fan
Pumped gas temperature is
too high
Add a cold trap to the intake port
Poor lubrication
Check oil level and filter
Pump Running
Hotter than Usual
The oil filter or exhaust line
is clogged
Replace the oil filter or clean exhaust line
Oil is unsuitable
Change the oil
Oil channel is obstructed
Clean the channel
Oil is insufficient
Add the oil to the level as requested
The condenser is dirty
Clean the condenser
Ambient temperature is too
high
Reduce ambient temperature
Oil comes from the vacuum
system
Check the vacuum system
Oil in the Intake
Line or in the
Vacuum Vessel
Anti-suck back valve spring
is obstructed
Replace anti- suck back valve spring
Anti-suck back valve plate
is obstructed
Replace anti- suck back valve plate
Oil level is high
Drain excess oil
Vacuum Pressure
in System Rises
too Fast when
Pump Stops
Vacuum system leak
Check the system
Anti-suck back valve is
malfunctioning
Repair anti-suck back
Too much oil in the pump
Drain some oil
Too Much Oil in
the Exhaust Port
Continuous operation under
high pressure in the intake
port
Shorten exhaust time
Oil mist filter is obstructed
Replace oil mist filter
Sealing Surface
Leak
Seal is damaged
Replace the seal
Seal Ring is damaged
Replace the seal ring