15
2) Connect the black banana plug of the AC/DC Special Current
Clamp Adapter to "
COM
" socket and the red banana plug to the
"
µAmA
" socket.
3) Press “
SELECT
” key to choose “
DC
” or “
AC
” measurement.
4) When perform DC current measurement, always rotate or push
the DCA zero adjuster on the Clamp Adapter until the multimeter
reads zero.
5) Clamp the Jaws around the
one
conductor to be measured.
Center the conductor within the Jaw using the Centering Marks as
guides.
6) Read the result from the LCD panel. The arrow in the Jaw
indicates the DC current direction of positive current flow (positive
to negative).
3.4 Resistance Measurement
1) Set the selector switch to desired “
” position.
2) Connect the black test lead to "
COM
" socket and red test lead
to the "
VΩHz
" socket.
3) Press “
SELECT
” key to choose
Resistance
measurement.
4) Connect tip of the test leads to the points where the value of the
resistance is needed.
5) Read the result from the LCD panel.
Note:
When take resistance value from a circuit system, make
sure the power is cut off and all capacitors need to be discharged.
3.5 Capacitance Measurement
1) Set the selector switch to desired “
” position.
2) Connect the black test lead to "
COM
" socket and red test lead
to the "
VΩHz
" socket.
3) Connect tip of the test leads to the points where the value of the
capacitance is needed.
4) Read the result from the LCD panel.