58
Device functions
RAMP
After every blower start, the device‘s microcontroller
checks whether a soft start ramp has been programmed.
The soft start function slowly increases the pressure
over the time programmed by you, starting with a pro-
grammable initial ramp pressure (P-Ramp) up to the
prescribed pressure (point 2 CPAP) or P-Start (point 2
AutoCPAP) in order to make it easier for you to get to
sleep. The soft start can bring relief particularly if you
are not quite used to the respiratory therapy yet.
The device offers delay time settings from zero (pre-
scribed therapy pressure starts immediately) to 30 min-
utes (prescribed therapy pressure is built up gradually
within 30 minutes). The initial ramp pressure (P-Ramp)
can be set between 4 hPa and the prescribed pressure.
Q),MMMMMMMMMMMM=;)%*
4%/!*1
Q ),MMMMMM? @$
4%/!*1