16
Operation
HP289
3.3
Calculation of Rate and Total
3.3.1
Frequency Input
The flowrate, R, is calculated as follows:
where
f
is the input frequency in Hz.
H
is the timebase of rate and is 1 for seconds, 60 for
minutes,3600 for hours, and 86,400 for days.
S
is the Scaling Factor.
The Scaling Factor,
S
, is equal to the K-factor of the flowmeter
expressed in pulses per unit volume.
The user sets the Scaling Factor and selects the timebase during the
configuration process as detailed in Section 5 of this manual.
S
fxH
R
=