Manufacturer’s guidelines and declarations — electromagnetic emissions
The A500LW09 compressor nebuliser is intended for use in the electromagnetic environment
described below.
The customer or the user of the A500LW09 compressor nebuliser should ensure its use in
such an environment.
Immunity tests
Test level
acc. to
IEC 60601
Compliance
level
Electromagnetic environment
— guidelines
Immunity to conducted
radio-frequency
disturbance fields
according to 61000-4-6
3 Vrms
150 kHz
to 80 MHz
3 Vrms
Portable and mobile radio
communication devices should
not be used near any part of the
nebuliser, including cables, at the
distance less than the recommended
distance calculated on the basis of an
appropriate equation adapted to the
frequency of the transmitter.
Recommended distance.
Radio-frequency
radiation according to
61000-4-3
3 V/m
80 MHz
do 2,5 GHz
3 V/m
80 MHz to 800 MHz
d
=[
P
]
7
E
1
800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitters
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres (m).
Field strength from fixed RF
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
m
27