background image

www.humboldtmfg.com • 1.800.544.7220

WW

Содержание PDGpro

Страница 1: ...www humboldtmfg com 1 800 544 7220WW ...

Страница 2: ...ame refer to Annex 1 11 Connect Devices 11 Measuring Procedure 12 Start Measurement 12 Quit Measurement 13 Error Menus 14 Evaluation Reading Printing Measured Data 15 Export the Stored Measured Data 15 Data Transfer Measuring Instrument USB Stick 15 Data Transfer Measuring Instrument PC 16 Delete Measuring Results 16 General 17 Settings Display 17 Device 18 Printer 18 Service 18 Calibration Menu 1...

Страница 3: ...s are highlighted as described below Warning This symbol is used in conjunction with related text to draw user s attention to hazards and risks which may cause failure of tester components or adversely affect operating procedures in case users do not take the corresponding precautions Note This symbol and the related text identify technical requirements and provide additional information to be tak...

Страница 4: ...con tained in this instruction manual and in the operating instructions from company LUKAS for the hydraulic components on supplied CD Compliance with the maintenance and servicing instructions con tained in this instruction manual and in the operating instructions from company LUKAS for the hydraulic components on supplied CD Any other use or any use beyond this definition is not intended and may...

Страница 5: ...r 2 Extension pieces 3 Hydraulic cylinder 4 Thrust piece 5 Load cell 6 Measuring tunnel 7 Load plate with bubble level 8 Hydraulic pump Settlement measuring mechanism measuring frame Figure 2 1 Extension arm with measuring finger 2 Traverse 3 Measuring frame 4 Measuring head 5 Dial gauge 6 Bubble level Electronic dial gauge Figure 3 1 Dial gauge 2 Dial gauge cable Figure 1 Figure 2 Figure 3 ...

Страница 6: ...c pump with 100 kN cylinder and 150 mm lift with high pressure hose of 2 m in length 1 set of hydraulic cylinder extension pieces plug type connection 1 pressure plate with magnetic holder and upper ball and socket joint Load plate with adjustable bubble level Diameter 300 mm Thickness of plate 25 mm Settlement measuring mechanism measuring frame Support frame on three point bearing equipped with ...

Страница 7: ... measuring instrument Measuring instrument data logger for load and settlement data with automatic processing and evaluation facilities as well as storage capability for more than 200 tests Comfortable user interface 3 5 TFT colour display Rechargeable Lithium ion polymer battery pack Lithium polymer battery pack Interfaces Bluetooth USB thermal printer GPS integrated thermal printer to print data...

Страница 8: ...elect to the right Scroll Confirm selection Start action Buttons Symbols The currently active button is displayed with colour the inactive buttons are grey Main menu Measuring page 11 Measured data page 15 read print export delete Settings Service page 17 display device printer support calibration General Back to the previous menu Discard frozen measuring values approach set point again Scroll tes...

Страница 9: ...of status line is the same in all menus Status Bluetooth displaying in status line in case Bluetooth is active during data transfer only Status GPS displaying in status line in case GPS is active and available State of charge of the printer displaying in status line in case printer is connected State of charge of the measuring instrument Overview Menu Functions Measuring Taking measurement y Measu...

Страница 10: ...es Switching off automatically The measuring instrument HMP PDGpro switches off automatically in case there is no action for about 90 s The device will not switch off automatically as long as it is in the measuring mode If the Lithium polymer battery pack of the measuring instrument drops below the voltage required for operation the device switches off automatically in order to prevent a deep disc...

Страница 11: ...connect it with the USB car charger via USB cable Connect the USB car charger to a car socket 12 V or via AC DC adapter to mains supply 230 V 50 Hz Disconnect the charger from mains supply when charging of Lithium polymer battery pack has been finished It is not possible to overload the Lithium polymer battery pack since it is equipped with an overload protection When the battery pack is fully cha...

Страница 12: ...tudinal beam of the measuring frame 3 and align with height adjustment star grips Remove the extension arm 1 from the bracket insert it into the measuring frame and fix it by means of the 2 star grips Insert the measuring finger into the measuring tunnel until the relevant mark is reached and align horizontally by using the circular bubble Place the dial gauge in the dial gauge bracket of the meas...

Страница 13: ... recording the GPS data then in the menu settings device Z has to be chosen for GPS page 18 Immediately after starting device the button Measuring is active and the measuring process can begin Before starting the measurement the desired measuring range plate diameter has to be chosen within the measuring instrument page 18 Confirm the button Measuring Figure 11 by pressing the key The menu Measuri...

Страница 14: ...15 The number of the current measuring series and the current measuring range the number of the current measuring point and also the set value for the pressing are displayed Accept the values after the required waiting time by pressing the key The values are frozen in the display Confirm button E by pressing the key to store the frozen measuring values for the current measuring point Figure 16 The...

Страница 15: ...trument provides instructions to monitor the measu ring procedure which pop up as an error report before measurement or during the measurement The following error reports might appear before the measurement Error report Error cause σ c error no connection between measuring instrument and measuring head σ error no connection between load cell and measuring head s error no connection between dial ga...

Страница 16: ...igure 21 By confirming the buttons M or N with the values of several mea suring points GPS position and settlement curves Figure 22 can be displayed Select button B and confirm with key Data of selected test series are printed GPS data will only be printed in case GPS is enabled J in menu Settings Device and when GPS data are stored with the selected measuring series Export the Stored Measured Dat...

Страница 17: ...and disconnect measuring instrument from PC To transfer data from measuring instrument to PC see instruction manual Protocol software for the Plate Load Tester HMP PDGpro Delete Measuring Results The test series and results stored in data base can be deleted via Button D Select and confirm the button in the main menu Confirm button D by pressing the key Figure 25 The menu on the left is displayed ...

Страница 18: ...e key Change the year by means of the keys Confirm the current date setting by pressing the key By pressing the key the menu date can be left at any time Select Footer G by means of the or key and confirm with key The menu Settings appears Select Footer G by means of the or key and confirm with key The set date will be saved and the main menu appears Set Time Select button Time and confirm with ke...

Страница 19: ...0 762 Kraft set measuring range of load plate or force GPS J Z activate deactivate GPS Einheit MN m MPa set unit Calib date J Z show don t show calibration date on start screen Printer In menu Printer B the following settings can be carried out for printer configuration Figure 30 Head data J Z print out protocol head always active Date time J Z print out date time Grafik J Z print out curve Servic...

Страница 20: ...30 V 50 Hz The AC DC adapter is included in the delivery contents of measuring instrument HMP PDGpro Charger and adapter are placed in the carrying case The printer AP1300 is shipped with a connected and fully loaded power pack When the printer is used for the first time after a lengthy period or has been standing idle for a lengthy period recharge the power pack prior to use In the event of malfu...

Страница 21: ...he printer should be recharged as quickly as possible when state of charge is 15 or less or in case the note on the left Figure 33 appears on the screen of measuring instrument Front panel of printer Figure 34 1 Paper feed Single line paper feed Press the key for a short interval and release Multi line paper feed Hold down this key until the desired length of paper is reached 2 LED Signals READY L...

Страница 22: ... is provided with sensors to detect lacking paper or opened paper compartment If a sensor is activated the printer switches to the storage mode all data transmitted to the printer are preserved Printing is continued immediately as soon as the defect has been removed It is recommended using original paper rolls for thermal printer only dimensions 3 cm width 5 7 cm length of paper 10 m Malfunction P...

Страница 23: ...HMP Magdeburger Prüfgerätebau GmbH has calibrated instrumentation used to conduct force and distance calibrations In addition necessary repairs may be carried out Hotline HMP Magdeburger Prüfgerätebau GmbH Bülstringer Straße 6 D 39126 Magdeburg Tel 03 91 2 51 46 66 03 91 2 51 46 67 Fax 03 91 2 51 46 68 E Mail info hmp online de ...

Страница 24: ...Name Benennung Datum Datum Bearb Gepr Buch stabe Zeichnungsnummer zuläss Abw für Masse ohne Toleranzangabe Halbzeug Werkstoff 2004 Änderung Masstab Hietschold Plate Load Tester PDG 1000 05 10 3 6 4 5 2 1 Nr Term 1 Extension arm with measuring finger 6 Hydraulic pump 5 Hydraulic cylinder 4 Load plate 3 Measuring frame Traverse 7 7 Measuring case ...

Страница 25: ...damages of the system The pump lever must be locked for transport Advancing cylinder Close the drain valve on the pump by turning the hand wheel clockwise to the right to advance the cylinder Unlock the pump lever and move it up and down to load pressure Also read the operating instructions of hand pump and hydraulic cylinder by LUKAS Company to get more information about advancing and retracting ...

Страница 26: ...on for the assessment of electrical and electronic products with respect to the restriction of hazardous substances EN 60335 2 29 2004 Household and similar electrical appliances Safety A2 2010 Part 2 29 Particular requirements for battery charqers EN 61000 6 2 2005 Electromagnetic compatibility EMC Part 6 2 Generic standards Immunity for industrial environments EN 61000 6 4 2007 Electromagnetic c...

Отзывы: