MAX14010_QuickSet_DE.doc 

Seite 3 von 3 

       09.05.2014 

 

Perform following steps: 
 
1. Press and hold button 1, turn on generator or reconnect the QuickSet, all LEDs light up, release button 
2. Setup parameters for button 1 at the generator 
3. Press and hold button 1 until all LEDs light up for one second 
4. Setup parameters for button 2 at the generator 
5. Press and hold button 2 until all LEDs light up for one second 
6. Turn off generator or disconnect QuickSet 
7. Press and hold button 2, turn on generator or reconnect the QuickSet, all LEDs light up, release button 
8. Setup parameters for button 3 at the generator 
9. Press and hold button 3 until all LEDs light up for one second 
10. Setup parameters for button 4 at the generator 
11. Press and hold button 3 until all LEDs light up for one second 
12. optional: Press buttons 1 to 4 shortly to check the correct operation of the generator 
13. Turn off and on generator or disconnect and reconnect QuickSet while not pressing any button 
 
From now on the QuickSet is prepared according the table above and is protected against accidental overwriting or 
changing the memory contents. 
 
 
 

6.  Trouble Shooting 

- QuickSet doesn't do anything, all LEDs remain dark all the time 
Be sure the generator ZZG5/ZZG6 is powered. 
Be sure you are using a straight through network cable, not a crossed cable. 
 
- While turning on the ZZG5/ZZG6 the LEDs flash but the settings cannot be sent to the ZZG5/ZZG6. After pressing 
and releasing a button the LEDs do not flash finally. 
Ensure the generator is equipped with firmware 1.26R. 
Ensure the GPS setting of the ZZG5/ZZG6 is "Sync msec: Off" 
 
- QuickSet doesn't store settings. After pressing and holding a button the LEDs do not light up for one second finally. 
Ensure the generator is equipped with firmware 1.26R. 
Ensure the GPS setting of the ZZG5/ZZG6 is "Sync msec: Off" 
Ensure you have pressed button 1 or button 2 while turning on the generator or connecting the QuickSet for entering 
one of the storage modes. 
 
- QuickSet seems to be working but the generator output voltage level always remains the same. 
Ensure the ZZG5/ZZG6 doesn't diplay the output voltage in dB-units. Turn off the dB-display (see above). 
 
 