HK Instruments DPT100 Скачать руководство пользователя страница 2

 

 

 

Technical data

 

Response Time

 

0.8 / 4 s selectable by jumper 

Max. pressure

 

25 kPa 

Bursting pressure

 

50 kPa 

Suitable media 

Air and non-aggressive gases 

Measuring element

 

Piezoresistive 

Electrical interface

 

Supply voltage 

24 VAC or VDC 

(3-wire) 

Max. tolerance 

±

 10% 

  

Power consumption 

< 1.0 W (<1.5W with Iout 20mA) 

Output signal 

  0…10 VDC, Load R minimum 1k

 or 

4…20 mA,  
maximum load 500

 

Materials 

Housing  

ABS 

Cover 

ABS 

Pressure connections 

ABS 

Duct connectors 

ABS 

Tubing 

PVC, soft 

Connections  

Electrical connections 

4 screw terminals, max 1.5 mm

2

 

Cable entry 

M16 

Pressure connections 

Male 

 5,0 mm and 6,3 mm 

Weight 

150 grams, with accessories 290 grams 

Dimensions 

90,0 x 71,5 x 36,0 mm 

   

 

General ambient condition 

Temperature range

 

Operation 

-10…+50

°

C (-5…+50

°

C for –AZ model)

 

Storage 

-20…+70

°

C

 

Ambient humidity 

0 to 95% RH

 

Safety

 

Protection standard 

IP54 

Conformance 

Meets the requirements for CE marking:  

 

EMC directive   89/336/EEC 
Rohs Directive   2002/95/EY 

 

 

Accessories

 

Standard accessories:

 

o

 

2 fixing screws 

o

 

2 plastic duct connectors 

o

 

2 m tube  

 4 / 7 mm 

 

Optional accessories: 

o

 

Metallic duct connectors 

Содержание DPT100

Страница 1: ...500 Pa 0 1000 Pa 1 5 1 Pa 8 Pa DPT2 5K D AZ Q 0 1500 Pa 0 2500 Pa 1 5 1 Pa 24 Pa DPT5K D AZ Q 0 2500 Pa 0 5000 Pa 1 5 1 Pa 24 Pa 4 range model R4 for 4 measuring range D for display AZ for autozero element Q for square root output range 1 range 2 range 3 range 4 Accuracy FS over operation temp 10 50 C range 1 4 Long term stability typical 1 year AZ without AZ DPT1000 R4 D AZ Q 250Pa 500 Pa 750 Pa ...

Страница 2: ...nnectors ABS Tubing PVC soft Connections Electrical connections 4 screw terminals max 1 5 mm2 Cable entry M16 Pressure connections Male 5 0 mm and 6 3 mm Weight 150 grams with accessories 290 grams Dimensions 90 0 x 71 5 x 36 0 mm General ambient condition Temperature range Operation 10 50 C 5 50 C for AZ model Storage 20 70 C Ambient humidity 0 to 95 RH Safety Protection standard IP54 Conformance...

Страница 3: ...the manual push button adjustment is not required Optional auto zero element Optional auto zero element makes the DPT transmitter maintenance free for periodical push button zeroing Element automatically adjusts the transmitters zero point from time to time this eliminates the zero point long term drift of the piezoresistive sensing element Zero point adjustment is carried out every 10 minutes Dur...

Страница 4: ...1000 The pressure values on the sticker depends on the model 500Pa 1000Pa 0 8s 4s jumper without jumper Jumpers 4 20mA ZERO 0 1 4s 500 1000 24VDC GND 0 10V Jumpers ZERO 0 8 4s Jumper 2 Jumper 3 4 range model e g DPT1000 R4 0 8s 4s 250Pa 500Pa 750Pa 1000Pa 250Pa 500Pa 750Pa 1000Pa The range sticker located inside the lid Pressure values depends on the model e g DPT1000 R4 Jumper 2 Jumper 3 Jumpers ...

Отзывы: