HK Instruments DPT Series Скачать руководство пользователя страница 3

Copyright HK Instruments 2018

www.hkinstruments.fi

Installation version 6.0 2018

STEP 4: SELECTING THE MEASUREMENT UNIT

1) Select the desired measurement unit. (see step 4)
2) Select the desired measurement range. (see step 5)
3) Select the desired response time. (see step 6)
4) Zero the device. (see step 7)
5) Connect the pressure tubes. Connect positive pressure to 

port labeled “+” and negative pressure to port “-”.

6) Close the lid. The device is now ready to be used.

 

STEP 3: CONFIGURATION

No jumper

Circuit open

Jumper stored

Jumper installed

Circuit closed

(Grey colour indicates that a jumper is installed)

Figure 3 - Jumper installation

2) Install jumpers on J1, J2 and J3 as required.
a. Using the range number from 1c above, find the corresponding 

range number in Chart 2.

b. Install jumpers on J1, J2 and J3 on device, as shown under the range 

number in Chart 2. (Grey colour indicates that a jumper is installed. 
Reference figure 3 for jumper installation.)

1) To change the measurement unit appearing on the display, install a 

jumper to both pins of J5 (see Figure 3).

2) Push down the zero button and the measurement unit options (Pa, 

kPa, mbar, inchWC, mmWC, psi) will cycle on the display.

3) To select a unit option to display, remove the jumper from J5 

while the measurement unit is visible on the display.

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Range 7

Range 8

Pa
kPa
mbar
inchWC

    0–25

0–0.025

     0–0.25

  0–0.10

      0–50

 0–0.05

     0–0.50

  0–0.20

  0–100

    0–0.1

    0–1.00

    0–0.40

  0–250

   0–0.25

  0–2.50

    0–1.00

-25–25

-0.025–0.025

-0.25–0.25
-0.10–0.10

-50–50

-0.05–0.05
-0.50–0.50

-0.20–0.20

-100–100

-0.1–0.1

-1.0–1.00

-0.40–0.40

-150–150

-0.15–0.15

-1.50–1.50

-0.60–0.60

Model DPT250-R8

Model DPT2500-R8

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Range 7

Range 8

Pa
kPa
mbar
inchWC

-100–100
-0.10–0.10
-1.00–1.00
-0.40–0.40

0–100

0–0.10

0–1.00

0–0.40

0–250
0–0.25
0–2.50
0–1.00 

0–500

0–0.50
0–5.00
0–2.00

0–1000

0–1.00
0–10.0
0–4.00

0–1500

0–1.50
0–15.0
0–6.00

0–2000

0–2.00
0–20.0
0–8.00

0–2500

 0–2.50

0–25.0

0–10.00

Model DPT7000-R8

Pa
kPa
mbar
inchWC

0–1000

0–1.00

0–10.0
0–4.00 

0–1500

0–1.50 
0–15.0
0–6.00

0–2000

0–2.0

0–20.0

0–8.00 

0–2500
0–2.50
0–25.0
0–10.0

0–3000
0–3.00
0–30.0

0–12.00

0–4000
0–4.00
0–40.0

0–16.00

0–5000
0–5.00
0–50.0

0–20.00

0–7000
0–7.00
0–70.0

0–28.00

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Range 7

Range 8

mmWC
psi

mmWC
psi

mmWC

psi

   0–2.6
0–0.0036

   0–5.1
0–0.0073

    0–10.2
  0–0.0145

    0–25.5
  0–0.0363

-2.6–2.6

-0.0036–0.0036

-5.1–5.1

-0.0073–0.0073

-10.2–10.2

-0.0145–0.0145

-15.3–15.3

-0.0036–0.0036

-10.2–10.2

-0.0145–0.0145

0–10.2

0–0.0145

0–25.5

0–0.0363

0–51.0

0–0.0725

0–102.0

0–0.1450

0–153.0

0–0.21725

0–204.0

0–0.2900

0–255.0

0–0.3625

0–102.0

0–0.1450 

0–153.0

0–0.21725

0–204.0 

0–0.2900 

0–255.0

0–0.3625

0–306.0

0–0.4350

0–408.0

0–0.5800

0–510.0

0–0.7250

0–714.0

0–1.0150

 

SELECTING THE MEASUREMENT RANGE 

CONTINUED

Chart 1

Range 7

Range 8

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Jumper J1
Jumper J2
Jumper J3

(Grey colour indicates that a jumper is installed. Reference Figur

e 3 and Schematics for jumper installation.)

Chart 2

The response time affects how fast the transmitter reacts to changes in 
the system. The response time is the time the device takes to reach 63 
% of the measured value. To smooth out unstable pressure fluctuations 
in airflow applications, select a longer response time.

Example:
Selected response time: 8.0 seconds
Result: Output signal achieves a new value in 20 seconds (Response 
time*5)

To change response time, install or remove jumper on J4.
(see Figure 3)
1) Install jumper on J4 for 8.0 second response time.
2) Remove jumper from J4 for 0.8 second response time.

STEP 6: SELECTING THE RESPONSE TIME

 

STEP 5: SELECTING THE MEASUREMENT RANGE

1) Determine the range number
a. Find the model in Chart 1.
b. Find the measurment unit (selected in step 4).
c. Find the required measurement range on the same line as the   

measurement unit (b above) and determine the range number in the 
header.

Содержание DPT Series

Страница 1: ...nitoring pressure and flow monitoring valve and damper control pressure monitoring in cleanrooms SPECIFICATIONS Performance Accuracy from applied pressure Models 250 and 2500 Pressure 125 Pa 1 2 Pa Pressure 125 Pa 1 1 Pa Model 7000 Pressure 125 Pa 1 5 2 Pa Pressure 125 Pa 1 5 1 Pa including general accuracy linearity hysteresis long term stability and repetition error Thermal effects Temperature c...

Страница 2: ...cation duct wall panel 2 Use the device as a template and mark the screw holes 3 Mount with appropriate screws Pressure sensor Zero button Connection for optional display Terminal block LED Jumpers YES NO NO Static pressure Filter Damper monitoring Fan Blower monitoring Not connected 90 0 77 0 71 5 53 0 4 3 95 0 36 0 STEP 2 WIRING DIAGRAMS For CE compliance a properly grounded shielding cable is r...

Страница 3: ...0 0 2 00 0 1000 0 1 00 0 10 0 0 4 00 0 1500 0 1 50 0 15 0 0 6 00 0 2000 0 2 00 0 20 0 0 8 00 0 2500 0 2 50 0 25 0 0 10 00 Model DPT7000 R8 Pa kPa mbar inchWC 0 1000 0 1 00 0 10 0 0 4 00 0 1500 0 1 50 0 15 0 0 6 00 0 2000 0 2 0 0 20 0 0 8 00 0 2500 0 2 50 0 25 0 0 10 0 0 3000 0 3 00 0 30 0 0 12 00 0 4000 0 4 00 0 40 0 0 16 00 0 5000 0 5 00 0 50 0 0 20 00 0 7000 0 7 00 0 70 0 0 28 00 Range 1 Range 2...

Страница 4: ...ransmitter have an autozero calibration See the product la bel If it shows AZ in the model number then you have the autozero calibration 1 Manual Pushbutton zero point calibration NOTE Supply voltage must be connected at least one hour prior to zero point adjustment a Disconnect both pressure tubes from the pressure ports labeled and b Push down the zero button until the LED light red turns on and...

Отзывы: