4.4

 

Installation 

The device can be used as a tabletop unit or installed in a device housing. 
At the bottom of the housing there are six M3 threaded holes for the in-housing installation. The 
maximum screw-in depth is 

4mm!

 When installing the device, please remove the rubber feet.  

 

 

 
 

4.5

 

Power Supply

 

The device requires an external DC voltage source of 24V

DC

 / min. 1.6A as the power supply. The 

connection is made via the 

X4 Power / Interlock

 connector at the 

+24V In

 and 

GND

 pins. 

 

Caution: 

Before you turn on the voltage source, please check its output voltage. Admissible range: 

+24V

DC

  ±10%. If the amplifier is operated with wrong supply voltage, it could become damaged. 

 
 

4.6

 

Ventilation 

Sufficient ventilation is required to remove the heat dissipated inside the case during operation (P

D

 of 

up to 36W). For tabletop operation, free flow of air is required. Active ventilation could be useful, 
depending on the working point and ambient temperature. 
If built into a device housing, the thermal losses can be removed through the structure of the housing, 
if its design allows for this. Otherwise, active ventilation is required. 
 
 

Содержание HA51U-3B2

Страница 1: ...1 hivolt de GmbH Co KG Oehleckerring 40 D 22419 Hamburg Germany 49 40 537122 0 49 40 537122 99 info hivolt de www hivolt de HA51U 3B2 High Voltage Amplifier 3000V Operating Manual...

Страница 2: ...eck After Delivery 9 4 2 Warning Notices 9 4 3 Temperature Compensation 9 4 4 Installation 10 4 5 Power Supply 10 4 6 Ventilation 10 4 7 Functional Test 11 4 8 Input Connection 11 4 9 Analog Control X...

Страница 3: ...function the device should immediately be put out of service and it should be secured against unintentional or accidental operation To provide protection for personnel in case of unit failure the safe...

Страница 4: ...protected against overload short circuit overtemperature transient overvoltage and high voltage flashover The LEDs on the front panel indicate operation and error statuses An isolated USB interface i...

Страница 5: ...functions The signal input is connected to a differential amplifier to avoid interference from ground loops It is available on SMB connector X2 and on X1 Analog Control It provides the analog setpoin...

Страница 6: ...pends on the load capacitance The effective load capacitance CL consists of the amplifier s internal output capacitance ca 300pF capacitance of the output line a typical coaxial cable ca 100pF m and c...

Страница 7: ...h voltage sources will be switched on The device will then still be in the Interlock shutdown state The high voltage generation is not functionally activated till the Interlock shutdown state is finis...

Страница 8: ...ansportation 20 25 50 70 C C Relative humidity Operation Storage and Transportation Not condensing 5 5 80 95 Depending on the ambient temperature modulation amplitude and load capacitance the maximum...

Страница 9: ...against any contact before the device is put into operation The test set up must be checked each time before the device is put into operation to ensure that it is not potentially dangerous It should...

Страница 10: ...connection is made via the X4 Power Interlock connector at the 24V In and GND pins Caution Before you turn on the voltage source please check its output voltage Admissible range 24VDC 10 If the amplif...

Страница 11: ...itch to HV on 7 The Power Enable and HV on LEDs will light up the other LEDs will not 8 Set the switch to HV off 9 The HV on LED will turn off 10 Disconnect the supply voltage 11 Open the interlock br...

Страница 12: ...3 I Digital input TTL level 12 GPI4 I Digital input TTL level 13 5VX O Auxiliary voltage output 5V max 10mA 14 GND Supply voltage GND 15 GND Supply voltage GND 16 IL_R O Interlock bridge to X1 IL_IN 1...

Страница 13: ...er no software download internal program update is possible 4 11 Power Interlock X4 Connector type Mini Combicon 5 pin Mating connector Phoenix Mini Combicon 5 pin FK MCP1 5_ 5 ST 3 81 Pin Signal Dire...

Страница 14: ...N S3 7 Must always be OFF OFF S3 8 Disable X1 Interlock ON Disable ON 4 15 LEDs LED Function Power on 24V supply voltage Overload on overload blink shutdown due to overload Overtemp on overtemperature...

Страница 15: ...functions The interlock circuit must be closed and an analog setpoint must be supplied via X1 or X2 High voltage is released by S1 Configuration control shutdown states reset and measuring functions...

Страница 16: ...N Parameter par par optional depending on the command Number format 1234 56 leading zeros decimal point and decimal places optional Parameter separator err 1N Error number cr End of record Carriage Re...

Страница 17: ...voltage IHVPAV A Actual value of positive HV supply current IHVNAV A Actual value of negative HV supply current VDC mV Actual value of supply HV DC DC converter 5V mV Actual value of 5 2V supply 15V...

Страница 18: ...Resistance 0 0 signal source output impedance 1 50 signal source output impedance default SV 3000 3000 Set Output Voltage Setpoint VONVD 10mV digital setpoint resolution 10mV Status word Bit Status b...

Страница 19: ...manually In the Device Manager entries are displayed under USB Controller USB Serial Converter and under Ports COM LPT USB Serial Port COMxx The baud rate is set to 115 kBd It cannot be changed Format...

Страница 20: ...not light up No supply voltage present HV on display will not light up S1 switch is in HV off position Interlock circuit is open X1 ENABLE signal not set to H Enabling by command EN is switched off O...

Страница 21: ...s Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU and comply with the following European standards EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and la...

Отзывы: