-5-
Q1E-EP1398
3. Operation
Procedure
1)
Confirm that the probe is cleaned and disinfected.
2)
Covering the probe with a disposal probe cover to protect the probe
is recommended. The probe cover should be allergy free material
to avoid allergic reaction. Apply acoustic coupling gel inside and
outside of the probe cover. Acoustic coupling gel is required as
a couplant.
3)
Connect the probe to the ultrasound diagnostic scanner, operate
the scanner, and adjust the image, all according to the instructions
given in the operation manual for the ultrasound diagnostic scanner
with which the probe is used as connected.
4)
Confirm the direction of the probe. The relationship between the
direction of the probe and the B-mode image is shown in Fig. . The
right-left orientation mark on the image indicates the direction
of the index mark on the probe.
5)
Perform the reprocessing procedure in accordance with the procedure
stated in “
4. Reprocessing procedure
” every time immediately after
completing the ultrasound examination.
6)
Store the probe in accordance with the storage condition indicated
in “
8. Specification
”.
Fig. 2 Relationship between Index Mark
and Right-left orientation mark
Right-left
orientation mark
Index mark
H