4.3 Troubleshooting
4 - 16
4.3 TROUBLESHOOTING
On occurrence of an abnormality, take a proper countermeasure
referring to Table 4-1.
Table 4-1 Troubleshooting Action Table
Instrument condition
Possible Cause
Action for Remedy
1 Nothing
displayed
on
screen even the power
switch is turned on.
Fuse blown out
Replace
2 Message
“Calibrate!”
appears on screen.
Wavelength calibration data
destroyed.
Calibrate instrument. Ask your
service section for remedy.
3
Message “D2 lamp1”,
“WI lamp!” appears on
screen
Lamp broken.
Replace
4
Error message appears
on screen
Improper operation
Redo following instruction
message on screen.
5 Large
fluctuation
in
displayed data
a. Lamp deterioration
b. Lamp contamination with
fingerprint
c. Dirt or water drip on cell or
window plate
a. Replace.
b, c. Remove.
If the instrument does not work normally even after the remedy
according to the above troubleshooting action table, contact your
local sales representative or service office of Hitachi
High-Technologies Corporation.
WARNING
This instrument has an electrical part inside that could cause
a dangerous electric shock when touched.
Leave inspection of inside of this instrument always to the
qualified service personnel.