Open   window   on   PC1   –   1,   this   scans   once   and   grabs   the   image   only   once   on   

PC1   –   1.         Or,   open   window   on   PC1-­‐

 

this   scans   continuously   (you   can   zoom   in   and   

out   and   choose   any   image   you   would   like   to   grab).   Once   you   ready,   click   on   red   Stop   
then   PC1   -­‐   

 

   grabs   only   one   image.   

   
   
Note:    When    button    A    (Run/Stop    section)    is    On,    means    image    is    live,    when    it    is   

Flashing   means   storing   image   in   SEM   only   one   at   a   time,   when   it   is   off   means   image   
has   been   stored   in   SEM.   

   

Taking   Pictures   with   Polaroid   
   

-

 

Focus   the   image   

-

 

Push   ABC   button   

-

 

Press   button   4   on   (Scan   Speed),   to   check   your   image.   

-

 

Press   button   A   on   (Run/Stop   section),   till   stops   flashing   and   off.         

-

 

Load   Film   

-

 

Lever   S/b   on   “L”   position   (to   the   right   

-

 

Insert   the   film   (Arrows   pointing   in)   

-

 

Facing   down   on   side   with   writing   (   this   side   facing   lens)   

-

 

Pull   film   cover   out   (   polaroid   remains   inside)   

-

 

Take   the   picture   by   pushing   Direct   button   

-

 

Push   the   film   cover   back   in      

-

 

Turn   lever   to   position   P   (left)   

-

 

Pull   hard   to   remove   film   

-

 

Wait   30   seconds   to   allow   to   develop   

-

 

Push   button   A   on   (Run/Stop   section)   

   
Shut   Down   Operation:   

   

-

 

Bring   magnification   to   40   

-

 

T   urn   Off   HV   (READY)   

-

 

Close   SC   AIR   Lock   Valve      

-

 

Turn   on   camera   

-

 

On   Joystick   push   Exch.   and   Enter   to   put   Specimen   Stage   on   Dismounting   position   

-

 

Open   Air   lock   lever   (   MV-­‐1)   to   transfer   sample   holder   from   SC   to   SEC   

-

 

Rotate   Transfer   Rod   C.W.   until   Rod   completely   in   

-

 

You   can   view   on   display   monitor   2   while   dismounting   from   stage   &transferring                  

(caution)   

-

 

Pull   back   the   Transfer   Rod   till   it   locks   

-

 

Close   Air   lock   lever   to   isolate   SC   from   SEC   

-

 

Turn   off   camera   

-

 

Put   gloves   on   

-

 

Push   Air   SEC   to   Vent      

Содержание S-4100

Страница 1: ...ation Hitachi field emission scanning electron microscope converted to electron beam writing by using nanometer pattern generating system NPGS Equipment Specifications Resolution 15 o A at 30 KV Magnification 20 to 300 000x Electron gun Cold cathode field emission electron gun ...

Страница 2: ...ves stay free of any kind of contamination to keep the inside system clean Samples Specimens must be completely dried Use carbon conductive tabs to mount sample on sample holder Adjust the height of sample holder with LEVEL GAUGE level gauge is on top of S E C chamber Loading Sample SC Air Lock Valve Closed Make sure High Voltage is OFF on display unit Set SC SEC switch to SEC MV 1 Valve is Closed...

Страница 3: ...sh FLASH STANDBY Push ON ON ADJ and keep eye on IP1 Emission current when IP1 pressure goes up degasses the tip Must Wait for Two Minute then TURN ON HI VOLTAGE Note The Emission Current should be 10µA if not push ON ON ADJ 3 Do ABC to adjust brightness and contrast 4 Go to TV Rate on SCAN SPEED section F or S 5 Need to Do Beam Alignment Go to COLUMN ADJUSTMENT or PF3 Key Board Chose Beam Align Us...

Страница 4: ... minimize movement on display Exit Go to COLUMN ADJUSTMENT or PF3 again Choose Stig Align Y Adjust with Y Stigmatism to minimize movement on display Exit 8 Focusing the Image use FOCUS KNOBS STIGMAN ALIGNMENT knobs Use COARSE focus then FINE focus Notice if image is stretching while you turn focus knobs in out of focus If so need to stigmate using the STIGMA ALIGNMENT knobs 9 Now Ready to operate ...

Страница 5: ...ight Insert the film Arrows pointing in Facing down on side with writing this side facing lens Pull film cover out polaroid remains inside Take the picture by pushing Direct button Push the film cover back in Turn lever to position P left Pull hard to remove film Wait 30 seconds to allow to develop Push button A on Run Stop section Shut Down Operation Bring magnification to 40 T urn Off HV READY C...

Страница 6: ... holder from Transfer Rod then close the door hold closed Push EVAC to Rough the SEC wait till green light for SEC VACUUM HI lights up Check the pressure for IP1 5x10 7 IP2 1 5x10 7 IP3 6x10 6 Turn off Display power ...

Отзывы: