5 Scope of delivery, transport, set-up and installation
5.5.3
Testing UV transparency
Once the pump cycle has been completed, check the vacuum chamber's
UV transparency as follows:
1.
Open the main measuring window in the spectrometer software.
2.
Select "Optic UV transparency test" from the menu item "Analysis".
Fig. 5-8
Testing the UV transparency
3.
Place the displayed (fig. 5-9) and freshly prepared recalibration sample on the
spark stand and fix it in place with the hold-down device.
4.
Switch on the excitation source at the rotary knob on the front of the instrument and
press "Start".
Fig. 5-9
Measuring the recalibration sample
Once the measurement is completed, the instrument will indicate the UV transparency
level.
If the message appears in
green
(UV purge factor min. 80 %), the spectrometer is ready
for operation.
Fig. 5-10
Argon purge completed successfully
If the message appears in
red
(UV purge factor < 80 %), please contact the service of
Hitachi High-Tech Analytical Science.
Содержание OE700 Series
Страница 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Страница 4: ......
Страница 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Страница 9: ......
Страница 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Страница 14: ......
Страница 20: ......
Страница 26: ......
Страница 36: ......
Страница 46: ......
Страница 48: ......
Страница 62: ......
Страница 66: ......
Страница 68: ......
Страница 70: ......
Страница 72: ......
Страница 74: ......
Страница 76: ......