5 Scope of delivery, transport, set-up and installation
►
22 Hitachi High-Tech Analytical Science
4.
Now start the application "Test Controls" (2, fig. 5-5, page 5-21).
5.
Actuate the "Start pump cycle" button (1, fig. 5-6).
•
The pump cycle starts and runs automatically.
•
Upon completion the message "Vacuum Pump cycle finished... Press OK to
continue..." will be displayed on the screen.
Fig. 5-6
"Test Controls" window
6.
Confirm the message with "OK" (1, fig. 5-7).
•
The pump cycle for establishing optical transparency has now been
completed.
Fig. 5-7
"Vacuum pump cycle finished" window
1
1