-9-
MN1-6268 Rev.2
4. Operation
This chapter describes the operation of the probe.
4-1. Insertion of the probe
Gently insert the tip of the probe into the insertion opening of the trocar.
4-2. Operation
Grasp the fin with a forceps and place the ultrasonic radiation part of the probe onto the inner organ surface during
surgery. An image of the region of interest is displayed on the monitor of the ultrasound diagnostic instrument. For
details on displaying and adjusting the image, refer to the documentation supplied with the ultrasound diagnostic
instrument.
Warning
Use an appropriate size trocar.
If an appropriate size trocar is not used, the probe may be damaged when inserted into the trocar. Use
a trocar with a diameter of 10 mm or more. Please refer to 3-4 for preparations before use.
Do not insert and operate the probe forcibly, as this may cause injury to a patient.
Insert the probe without excessive force on the probe and the cable. If the probe cannot be inserted
smoothly, stop using the probe.
Do not grasp the cable with forceps.
The probe may malfunction due to cable disconnection. Please operate by grasping the fin with forceps
always.