- 12 -
Q1E-EP1441
7.3
Cleaning Process
Prior to liquid chemical sterilization/high level
disinfection process, the probe must be cleaned.
7.3.1
Cleaning
Immerse the probe in an enzymatic detergent (Enzol
®
)
solution (as shown in Fig. 3) for more than 30 minutes
or until the visible residue is removed.
If needed, scrub the probe using a soft sponge, gauze,
or cloth to remove visible residue from the surface of
the probe.
Follow the instructions of the enzymatic detergent manufacturer
regarding recommended dilution, temperature and soak time. Using
insufficiently diluted detergent may damage the probe.
7.3.2
Rinsing and Drying after Cleaning
After cleaning and prior to liquid chemical
sterilization/high level disinfection, the probe must
be thoroughly rinsed with deionized running water or
sterile water for more than 10 minutes to remove all
residue of the detergent solution (See Fig.4). After
rinsing the probe, air-dry it completely.
Remove all debris and residue of sterilant/detergent solution from
the surface of the probe sufficiently.
Debris and residue of sterilant/detergent solution on the surface
of the probe may affect patient health.
WARNING
WARNING
DM 120918 v2