-12- Q1E-EP1191
4.4
Drying
1)
Wipe the probe with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the equipment.
2)
Dry the probe naturally in an ambient temperature between
15-30°C for a minimum of 4 hours. Alternatively the probe can
be dried using a drying heater at a temperature of less than
40°C.
4.5
Inspection
Inspect the probe for any damage such as crack, scratch or
deformation. Do not use it if any damage is found.
4.6
Storage
Store the probe in a cool, dustproof and dark space to avoid high
temperature, humidity and direct sunlight. Limitations for the
time for sterilized equipment belong to package.
1) Do not sterilize this probe by pressurizing/ depressurizing
sterilization system such as ETO and Plasma system. If you sterilize
the probe by these systems, the membrane that seals the oil bursts
and the probe will not be functional.
2) Do not sterilize this probe by autoclave. If you autoclave the probe,
the probe suffers serious damage and will be not functional.
Drying
WARNING