-12-
Q1E-EP0415
7.
Specifications
7.1.
Probe
Type: EUP-F531
Center frequency:
6.5MHz
Technology:
High density Convex Array Probe
Dimensions:
See Fig. 7
Weight: Approx.
0.6kg (incl. cable and connector)
Scanning angle:
100°
Probe materials:
Bio-compatible allergy free components
Acoustic output:
According to IEC60601-2-37 (See Main Unit
manual.)
Applicable system:
Depending on production and upgrade
status. For detailed information contact
a service support.
Classification:
MDD classification IIa.
Cleaning:
Applicable detergents are listed in the
suppliers list
Disinfection:
Applicable disinfectants are listed in
the suppliers list
Sterilization:
ETO gas sterilization
Operating conditions:
Ambient temperature;
25 – 35°C
Contact surface temperature
(temperature of examinee);
Max. 42°C
Relative humidity;
30 – 85%
Storage conditions:
Temperature;
–10 – +55°C
Relative humidity;
10 – 95% (subject to no condensation)