-3-

MN1-6119 Rev. 3

Unit: mm

2. Specifications and Parts name

The specifications and the name of each part are provided below.

2-1. Specifications

Application:

Intraoperative diagnosis

Type of patient contact:

Intraoperative

Connectable instruments:

ARIETTA 70, ARIETTA 60, Noblus, ARIETTA Precision
NOTE:

At the time of publication of this manual, the connectable diagnostic ultrasound 

instrument or instrument software version available with this probe is different 

for each country, please refer to the instrument instruction manual or contact 

your local Hitachi representative.

Field of view:

82°

Frequency:

3.5 MHz

Cable length:

2.9 m

Service life:

3 years

Applied part:

Probe tip including ultrasonic radiation part, see the section 2-2

Part treated as applied part:

Cable up to 0.2 m length from the probe tip

IPX7 range:

See Figure 1 (In case that not putting the waterproof case to the ultrasound 

probe connector) 

In case that putting the waterproof case to the ultrasound probe connector, 

whole the probe from the tip of the ultrasound probe to the connector with 

Waterproof Case WP-001 is IPX7. range  

Measurement accuracy:

Refer to the instruction manual of the ultrasound diagnostic instrument

External dimensions:

See Figure 1

Remark: The tolerance for the dimensions is ±10%.

Figure 1 External View 

 

 

1000

41

23

22

Cable length

IPX7 range

Applied part

Содержание C22I

Страница 1: ...nnel Please read through this Instruction Manual as well as the separate Instruction Manual Safety MN1 5984 and Cleaning Disinfection and Sterilization MN1 6000 carefully prior to use Keep this Instruction Manual together with the ultrasound diagnostic instru ment for any future reference Hitachi Ltd 2015 2017 All rights reserved ...

Страница 2: ...ii MN1 6119 Rev 3 ...

Страница 3: ...prevent hazards and injuries to the operator or patients Danger Indicates an imminently hazardous situation which if not avoided will result in death or serious injury to the operator or patient Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury to the operator or patient Caution Indicates a potentially hazardous situation which if not ...

Страница 4: ... 1 1 1 Intended use 1 1 2 Classification of ME equipment 1 1 3 Standard components 1 1 4 Option 1 2 Specifications and Parts name 3 2 1 Specifications 3 2 2 Name of each parts 4 3 Preparations before use 5 3 1 Visual check 5 3 2 Confirmation of cleaning disinfection and sterilization 5 3 3 Operation check 5 4 Operation 7 4 1 Operation 7 ...

Страница 5: ...range of IPX7 Classification based on the degree of protection against electric shock Type BF applied part Classification for protection against ingress of liquids IPX7 Watertight equipment Operation mode Continuous operation Method of sterilization Refer to the separate instruction manual Cleaning Disinfection and Sterilization 1 3 Standard components The standard components of C22I probe are as ...

Страница 6: ... 2 MN1 6119 Rev 3 ...

Страница 7: ...ve Field of view 82 Frequency 3 5 MHz Cable length 2 9 m Service life 3 years Applied part Probe tip including ultrasonic radiation part see the section 2 2 Part treated as applied part Cable up to 0 2 m length from the probe tip IPX7 range See Figure 1 In case that not putting the waterproof case to the ultrasound probe connector In case that putting the waterproof case to the ultrasound probe co...

Страница 8: ... corresponds to the side of the orientation mark on the image Probe tip This part is gripped during operation Cable Cable transfers electric input output signals Connector The connector is the part which is connected to the ultrasound diagnostic instrument Caution Do not pull bend twist or apply excessive force to the cable The probe may malfunction due to cable disconnection Do not subject the ul...

Страница 9: ...eration check Connect the probe to the ultrasound diagnostic instrument and check that the displayed scan type and frequency correspond to those of the probe Check also that there is no abnormality in the image Remark Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to connect the probe and information displayed on the monitor If the probe is operated in...

Страница 10: ... 6 MN1 6119 Rev 3 ...

Страница 11: ... Do not operate the probe with excessive force Use with excessive force could result in injury to the patient Scan for minimum time necessary at the lowest possible acoustic output Acoustic output may affect the patient s internal tissues For details about the acoustic output please refer to the documentation supplied with the ultrasound diagnostic instrument Do not touch the connector terminal pi...

Страница 12: ...08 31 Manufacturer Hitachi Ltd 2 16 1 Higashi Ueno Taito ku Tokyo 110 0015 Japan Contact 81 3 6284 3668 http www hitachi com businesses healthcare index html Overseas Offices Hitachi Medical Systems GmbH Otto von Guericke Ring 3 D 65205 Wiesbaden Germany EU Importer Hitachi Medical Systems Europe Holding AG Address Sumpfstrasse 13 CH 6300 Zug Switzerland ...

Отзывы: