Turn on the AP 720S. After initializing, the [Main Menu] screen 

will be displayed.

21.

Select the panel type (Panel 20 or Panel 36). Press “Next” and 

move to the [Check] screen.

22.

Ensure there is sufficient Antibody Reagent in the Reagent 

Reservoir, sufficient Wash Buffer in the Wash Buffer Bottle, and 

sufficient Deionized Water in the Deionized Water Bottle, and 

check that the Waste Bottle is empty.

23.

Press “Start” to begin automated processing. (There are 8 

processes in all to complete the Auto-Run).

Load Photoreagent Mixture

24.

At the end of Process #7, the [Change Photo Rgt] will appear, 

the AP 720S will pause and an alarm will sound. Press “Alarm Off”.

25.

Prepare Photoreagent Mixture by mixing an equal amount of 

Photoreagent AB and CD. Note: Allow Photoreagent AB and CD 

to come to room temperature prior to mixing. Once mixed, the 

Photoreagent should be used immediately for best results.

26.

Fill the Photoreagent Reservoir (RED line) with the required 

amount of Photoreagent Mixture as listed in the Instruction

Manual.

27.

Open the door. Remove the Antibody Reservoir and load the 

Photoreagent Reservoir in the Sample / Reagent Rack. Close

the door.

Continue Processing

28.

Press “Continue” on the [Change Photo Rgt] screen.

29.

Press “Yes” on the [Check] screen to start Process #8 

(Photoreagent Aspiration).

30.

As soon as the first pette is filled with Photoreagent Mixture, 

set a timer for 10 minutes.

31.

At the end of Process #8, an alarm will sound. Press “Alarm Off” 

on [Process End] screen.

Unload and Read Pettes

32.

Open the door, remove the Pette Rack from the instrument and 

set on the Pette Stand.

33.

Insert a black top plug into each Pette.

34.

Unload the pettes one by one from the Pette Rack.

 

35.

Remove the Pette Tip from each Pette.

36.

Insert a white bottom plug into each Pette.

37.

Load the Pettes in order into the Pette Cassette Tray.

38.

When the timer reads 10 minutes load the tray in the CLA-1 

Luminometer and read the test results.

Completion of Operation

39.

Remove the Sample / Reagent Rack and discard all used sample 

cups.

40.

To run a second set of pettes immediately following the first set, 

press “End” on the [Process End] screen to return to the [Main 

Menu] screen.

41.

Alternatively, to prepare for shutdown, press “Rinse” on the 

[Process End] screen.

42.

Unscrew the caps on the Wash Buffer and Deionized Water 

bottles. Pull out the tube from the Wash Buffer Bottle and place 

it in the Deionized Water Bottle along with the Deionized Water

tube. Press “Ok” in the [Rinse] screen.

43.

On completion of the rinse remove the Wash Buffer and Deionized 

Water tubes from the Deionized Water bottle. Place Wash Buffer 

and Deionized Water tube on a clean surface. Press “Ok” on the 

[Drain] screen.

44.

After the drain function is finished the [Main Menu] screen 

appears. Turn off instrument.

45.

Rinse Reagent Reservoirs with deionized water.

Doc. No. 0822 - ENG Rev. 00

www.hcdiagnostics.com

QUICK START

TM

AP 720S   Semi-Automated Instrument (continued)

[Main Menu]

[Panel  Type]

[Check]

[Change Photo Rgt]

Elapsed  Time  0:0:15

[Check]

Add Photo Rgt?

[Process  End]

Elapsed  Time 0:0:15

Antibody,

Wash Buffer,

Water.Waste

[Panel 36]

Assay

Panel  36

Next

Back

Start

Alarm Off

Yes

Alarm Off

Rinse

End

No

Continue

[Rinse]

Please  change  tube

to Water  Bottle.

[Drain]

Please  pulll  out  tube

from Water  Bottle.

Back

Ok

End

Ok

36

20

Maintenance

07/ 2/23

14:25:21