Hisense Babysense 7 Скачать руководство пользователя страница 4

3

See illustra�on 7 for layout of control unit featuring:

• On/Off bu�on.

• Green breathing movement indicator light.

• Acous�c and visual red alarm light.

• Low ba�ery red warning light.

• Yellow shutdown indicator light

Babysense

 is ready for opera�on and use once the 

installa�on is completed. To operate, follow the below 

instruc�ons:

• Place your baby in crib/bed.

• Turn control unit on by pressing On / Off bu�on.

• Make sure red control lights switch on briefly and a 

DOUBLE click sound is heard. 

• When breathing and movement are sensed, the green   

light flashes. 

• Acous�c and visual red alarm light are ac�vated if 

baby's breathing stops or slows significantly.

• If red low-ba�ery warning light flashes, replace all 

ba�eries as soon as possible. If in addi�on clicking is 

heard, all ba�eries must be replaced immediately. 

• Turn off by pressing On / Off bu�on before taking 

your baby out of crib / bed. Make sure the Yellow 

shutdown indicator light momentarily blinks and no 

other lights are seen therea�er to indicate the device 

is turned off.

Babysense 

is a highly sensi�ve device, which monitors 

baby's breathing movements through the ma�ress 

during sleep. There are two main causes of malfunc�on - 

technical malfunc�ons and malfunc�ons due to 

environmental factors.

 

In order to minimize technical problems, follow 

installa�on and opera�on instruc�ons carefully.

Due to the high sensi�vity of Babysense, fans, wind, air 

condi�oning, and so forth can influence it. Such 

environmental effects could be picked up by the sensors 

on a similar level to that of breathing movements, thus 

preven�ng or delaying alarm ac�va�on.

If the Babysense is concurrently used with other devices, 

it is required to conduct the daily performance test 

(below) under an environment iden�cal to the actual 

situa�on and while all concurrently used devices are 

"ON". Under such condi�ons, it is required to verify that 

the acous�c and visual alarms of Babysense are ac�vated. 

 

Do not conduct performance test by picking up baby 

out of bed and wai�ng for the alarm to sound   

• The performance test should be conducted while baby 

is not in bed. 

• Turn device on and take two steps away from bed. 

• Ensure that alarm goes off within 20 seconds (due to 

absence of breathing movements).

• The alarm may sound with or without preliminary �cking. 

• Check that alarm can be heard in all loca�ons 

throughout the home or place of care.

• In case that alarm does not sound within 20 seconds 

refer to troubleshoo�ng on page 4-5.

• Performance test must be conducted daily and every 

�me you move the monitor, change bedding or 

relocate crib/bed posi�on.

 

Isolated cases of false alarms are possible. Check if 

caused by faulty connec�on of the sensor pad(s) to the 

control unit, connec�on of only one sensor pad when 

baby can move off the immediate area or failure of the 

sensor pads to be directly against the ma�ress.

In case of false alarms, kindly refer to troubleshoo�ng 

guide on page 4-5. If the problem cannot be solved, 

please contact manufacturer’s service personnel.

 

Act immediately according to the following: 

1. 

Approach the bed and check that baby is breathing. 

Note pallor (color of skin), muscle tone and movement.

2.

 If baby does not seem normal, pinch baby's heel and 

check that there is no cause for choking, such as an 

object lodged in baby’s throat. 

When breathing resumes, consult a doctor.

3.

 If breathing does not resume, apply first aid and seek 

medical assistance immediately.

Layout of control unit

Opera�on

Tes�ng Opera�on

Technical malfunc�on

Environmentally caused malfunc�on

Conduct this performance test daily

False alarms

In the event of alarm

I

llustra�on 7

Green breathing 

movement 

indicator light

Red

alarm light

Red low ba�ery 

warning light

Hidden On/O

ff

power bu�on

3

User Guide

Yellow shutdown

indicator light

Отзывы: