41
Scan Operation
4.4 Scan Operation
Scanning is started by the SCAN signal of EXT. I/O or the communication
*
TRG
command. The scan
channel is moved forward by the SCAN signal or the
*
TRG
command.
When the channel is the final channel, scanning is completed by the SCAN signal or the
*
TRG
command,
and all relay is opened and goes back to the beginning of the scan list. Note that scan operation will not
complete until another SCAN signal or
*
TRG
command is input in the final channel state.
Scan operation when the trigger source is set to STEP
SCAN signal or
*
TRG
command
Channel
Open
Start
channel
Open
Next
channel
…
Open
Final
channel
Open
Scan complete
CLOSE signal
OFF
ON
The CLOSE signal is output in pulses. The pulse width can be set using the following command
:
IO:PULSe:TIME <0.001
to
0.100/MIN/MAX/DEF>
If the next channel is closed before the CLOSE signal pulse reaches the set pulse width, CLOSE
signal pulse will automatically turn off.
Item
Communication command
Setting
procedure
*TRG
Starts scanning and proceeds along the scan channels.
:ABORt
Aborts scan operation and returns to the beginning of the scan
channel.
Setting
example
Executes scanning twice from CH1 of
SLOT 1 to
CH22 of
SLOT 3
.
:SCAN 101:322
:TRIG:SOUR STEP
Advances the scan channel one step at a time using the
*TRG
command (or SCAN signal).
*TRG
(Repeats for the number of channels.)
*TRG
Scan operation complete by final channel state.
Scanning is performed again after completion.
*TRG
(Repeats for the number of channels.)
*TRG
Scan operation complete by final channel state.
Aborts scan operation and restarts from the beginning.
:SCAN 101:322
:TRIG:SOUR STEP
Advances the scan channel one step at a time using the
*TRG
command (or SCAN signal).
*TRG
(Repeats.)
:ABOR
(Aborts scan operation and returns to the beginning of the scan
channel.)
*TRG
(Repeats for the number of channels from the first channel.)
All channels are open after scanning is stopped.
The following commands are disabled during scanning.
•
Setting the connection method
•
Setting the shield switching
•
Channel switching
•
Setting the channel delay
•
Setting the scan list
•
Setting concerning the EXT. I/O
•
Channel control due to CLOSE command
4
Scan Function
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...