37
Channel Delay Function
3.7 Channel Delay Function
The delay time after channel switching can be specified. When the specified
delay time elapses
after the channel relay is closed, a CLOSE signal of EXT. I/O is output.
If the measurement response time for the instrument needs to be ensured, set the delay time.
The required delay time depends on the instrument to be used and the measuring object.
Channel delay time setting example
Model
Delay time
BT4560
1 ms
BT3562A
10 ms
3561
3 ms
DM7276
0 ms
The channel delay time can be specified for each slot. The same delay time is used for all channels
within the slot.
For the switching sequence, see "3.4 Channel Switching Operation" (p. 31)
Item
Communication command
Setting
procedure
:SYSTem:MODule:DELay <
slot No.
>,<
delay time
0
to
9.999[
s
]>
Setting
example
Set the channel delay time for
SLOT 1
to 0.01 sec. and for
SLOT 2
to 0 sec.
:SYST:MOD:DEL 1,0.01
:SYST:MOD:DEL 2,0
3
Channel Switching
HIOKI SW1001A961-04
Содержание SW9001
Страница 2: ...HIOKI SW1001A961 04...
Страница 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Страница 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Страница 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Страница 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Страница 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Страница 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Страница 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Страница 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Страница 164: ...160 HIOKI SW1001A961 04...
Страница 167: ...HIOKI SW1001A961 04...
Страница 168: ...HIOKI SW1001A961 04...
Страница 169: ...HIOKI SW1001A961 04...